Re: [AMMRL] Flirting with the 5 gauss line #HeliumRecovery

From: Fulton, Luke <fko24003_at_uconn.edu>
Date: Fri, 21 Jun 2024 17:41:07 +0000

Hi all,

Thanks for the feedback! We'll probably use Bob's suggestion to stagger the
installation while waiting for our new magnet. I was fantasizing of some way
to avoid reconfiguring the plumbing connections to the storage tanks. Alas,
can't have that cake and eat it too.

Those compass phone apps work great, much more sensitive than expected. With
how quickly the field climbs while approaching the magnet I'm realizing we
don't anything to do with it while moving tanks. There's limited space to
maneuver and the tanks will be set against a wall. It's a conceivable a person
might step between the tank and magnet while positioning, and if the field
grabbed hold while they were there...shudder.

Kind regards,
Luke

Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279

email: fko24003_at_uconn.edu
(alias): luke.fulton_at_uconn.edu
mobile: (603) 953-5275
Office: (860) 486-4069
________________________________
> From: main_at_ammrl.groups.io on behalf of Sameer Al-Abdul-Wahid via groups.io
> Sent: Friday, June 21, 2024 12:23 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Flirting with the 5 gauss line #HeliumRecovery

Hi Luke,

Our unshielded 400 MHz is ~7.8 ft to the 5G line, and ~6.2 ft to the 10 G line,
as per the chart in our magnet manual. If that’s not available, you
might consider trying a smartphone app to confirm. I don’t know if
that helps – but maybe an engineer would sign off if it was “only” 10 G?

If you do go ahead, one concern is metal shavings from drill bits. For our
recovery system, we had to drill an aluminum support post just near the 5 G
line of one of our magnets. We had a vacuum running beside the drill bit to
capture dirt and shavings. As a backup, we made a ‘wall’
around the drill site with cardboard boxes.

Best,

Sameer

--

Sameer Al-Abdul-Wahid, Ph.D. | Manager
Nuclear Magnetic Resonance Centre | University of Guelph
519-824-4120 ext. 58914 | s.wahid_at_uoguelph.ca<mailto:s.wahid_at_uoguelph.ca>
http://nmr.uoguelph.ca/


> From: main_at_ammrl.groups.io on behalf of Bob Berno
> Date: Friday, June 21, 2024 at 11:05
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Flirting with the 5 gauss line #HeliumRecovery

Hello Luke,

You say that the old 400MHz magnet is “… not decommissioned yet.”
Implicit in that statement is that the old 400MHz magnet will be decommissioned soon.

My suggestion: put a stop to the installation of the storage tank in question
until the old magnet is de-energized. Could your recovery system operate with
two storage tanks temporarily? Then, once the old magnet is removed, the third
storage tank added?

Too many things can go horribly wrong when you have a big stray field and large,
heavy magnetic objects, as well was people working with heavy metal tools.

And this doesn’t even consider how difficult/impossible it will be
to record any useful data using the 400MHz.

Good luck.

Bob.

> From: main_at_ammrl.groups.io On Behalf Of Fulton, Luke via groups.io
> Sent: June 21, 2024 10:45 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Flirting with the 5 gauss line #HeliumRecovery

Hi All,

I'm seeking the collective wisdom regarding a site planning pickle.

We're replacing our old 400 MHz magnet/console later this summer. The grant
also included a full helium recovery system with three medium pressure storage
tanks that we're in the midst of installing. I think it was assumed the new
magnet would be installed before the recovery system, and that didn't happen.
The old magnet hasn't been decommissioned yet and it's unshielded. The 5 gauss
line extends ~7ft from the magnet, and the plan has one of the tanks crossing
it by a full foot. Each tank is 8.5ft ht, 3ft dm, magnetically susceptible
(steel?), and ~400 lbs. We have to permanently bolt down all three next week.
I can rearrange the plan, but it would be a tragedy for the lab layout, and
I'd never hear the end of it.

Does it seem like an acceptable risk to put that much metal inside the stray
field?

My building manager thinks we could slide the magnet a foot or two out of the
way with johnson bars. I've only ever heard the direst of warnings about trying
to move a magnet, and I'm skeptical we could even pull it off if we tried.

Kind regards,
Luke

Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279

email: fko24003_at_uconn.edu<mailto:fko24003_at_uconn.edu>
(alias): luke.fulton_at_uconn.edu<mailto:luke.fulton_at_uconn.edu>
mobile: (603) 953-5275
Office: (860) 486-4069




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1365): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1365__;!!PvDODwlR4mBZyAb0!V77fGZfF1mSgS8CrK8mw1i4Ubkw5_2=
v763281cb2rYAcaLXMtP6tffpi0DpYaKjB2l2kOVqyt3XFA6LPZA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106800203=
/7559972__;!!PvDODwlR4mBZyAb0!V77fGZfF1mSgS8CrK8mw1i4Ubkw5_2v763281cb2rYAca=
LXMtP6tffpi0DpYaKjB2l2kOVqyt3UeJsNytg$
Mute #heliumrecovery:https://urldefense.com/v3/__https://ammrl.groups.io/g/=
main/mutehashtag/heliumrecovery__;!!PvDODwlR4mBZyAb0!V77fGZfF1mSgS8CrK8mw1i=
4Ubkw5_2v763281cb2rYAcaLXMtP6tffpi0DpYaKjB2l2kOVqyt3W2TrV1ew$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 21 2024 - 10:41:13 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST