Our Avance 600MHz spectrometer started to give errors on acquisition commands.
The system locks, but fails on zg , wobb, and ii commands with various
aqrackserver exceptions. The +9V LED on the PSM3 is not lit. I tried replacing
fuse 4 in the PSM3 module, but the LED is still off. I have tried restarting the
console, doing cf, ii, etc but the errors are still there. The PSM3 on this
spectrometer had failed before (predating my time here). Is this a failure of
the PSM3 module or is something else going on, like a communication issue?
The error messages we received are as follows:

1. On executing zg command>

Error during BirdRack flush:
Error in flush():
aqrackserver exception:
Error in "flush": Cannot disable observe for Sgu: Error 140 in RX_1:BH SLOT
2 RACK 1:Power diagnostic failed

2. On executing wobb command>

Warning: cannot put RXS-1 asleep:
(Error in sleep: Error in sleepRXS():
aqrackserver exception:
sbsb error in "sleepRXS": Error 140 in RX_1: AW SLOT 2 RACK 1: Power
diagnostic failed) Wobble curve may show spikes

3. On executing ii command>

Error during BirdRack flush:
Error in flush():
aqrackserver exception:
sbsb error in "flush": Error 211 in ROU1: BH SLOT 8 RACK 1: No l2c
acknowledge received

I would greatly appreciate it if the AMMRL community could provide some
guidance about this. Also, if it is a PSM3 issue, is anybody willing
to part with a spare they might have?

Thanks,
~Suv

Suvrajit Sengupta
Director of NMR Facilities
University of California, Irvine
Irvine, CA 92697, USA


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1346): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1346__;!!PvDODwlR4mBZyAb0!UOMxIXEWettp2U5Adn0BfXAygUsn7f=
BKl5CIY5oDGZ6kilLF5XVsZJBT4UWV0PpBWFiIeDCqYJJD52vx$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106750944=
/7559972__;!!PvDODwlR4mBZyAb0!UOMxIXEWettp2U5Adn0BfXAygUsn7fBKl5CIY5oDGZ6ki=
lLF5XVsZJBT4UWV0PpBWFiIeDCqYOHU5Xci$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!UOMxIXEWettp2U5Adn0BfXAygUsn7fBKl5CIY5=
oDGZ6kilLF5XVsZJBT4UWV0PpBWFiIeDCqYPmjqFsk$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!UOMxIXEWettp2U5Adn0BfXAygUsn7fBKl5=
CIY5oDGZ6kilLF5XVsZJBT4UWV0PpBWFiIeDCqYMZatgrT$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jun 18 2024 - 15:41:21 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST