Dear Emeric,
you can find TTL pulses on the IPSO unit. In the T-controller section. The
best of the 4 blocks to access is W
Somewhere in your accessories box there should be the matching 6 pack connector
with 6 cables with BNC connectors at the end. The pins in one of these connectors
should be numbered 1 - 6
If we use connector W we would deal with W1 - W6.
These can be addressed by commands in the pulse program with the setnmr(word)|
or ^(bit) command. The pins are mapped to NMR control words. On connector W you have
W2 = word 3, bit 12
W4 = word 3 bit 13
W6 = word 3 bit 14
W1 = word 4 bit 7
W3 = word 4 bit 8
W5 = word 4 bit 9
Examples are now setnmr3^12 set W2 high or setnmr3|12 set W2 low.
You can then use define statements to make the commands more userfriendly.
Create devices.incl file in the pp/user folder and add #include <devices.incl>
#define Device_ON setnmr3|12 if you use active low logic. the use Device_ON in the pulse program.
I hope this helps
Best regards
Clemens
On 7/21/2026 9:19 AM, Wasielewski Emeric, Ph. D. via groups.io wrote:
Hi everyone,
We would like to control some external electronics attached to a Bruker
Avance III HD nanobay spectrometer, running Topspin 3.8 software.
What we are looking for is TTL signals that can be turned on and off during
the pulse sequence.
Anyone out there that has accomplished this please let us know how, or where
to find more information than in the classic Bruker books.
Thanks and regards,
Emeric
--
Emeric Wasielewski, PhD
Responsable Plateforme RMN Cronenbourg
Tel : +33 (0)6 10 87 22 93
Courriel : wasielewski_at_unistra.fr
Laboratoire d'Innovation Moléculaire et Applications
UMR CNRS 7042-LIMA
Univ. de Strasbourg | Univ. de Haute-Alsace
ECPM
25 rue Becquerel
FR-67087 Strasbourg 2
lima.unistra.fr <
http://lima.unistra.fr/>
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3601): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3601__;!!PvDODwlR4mBZyAb0!X268eVdHltPkTi_Jgc-SnpF6OuOLmy=
N7W0erVM4LyQxdmypw7uWxkuh-LYN_Y87tFQojDF8YWA5x5h-XSBi33Kk7nOsN$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120376618=
/7559972__;!!PvDODwlR4mBZyAb0!X268eVdHltPkTi_Jgc-SnpF6OuOLmyN7W0erVM4LyQxdm=
ypw7uWxkuh-LYN_Y87tFQojDF8YWA5x5h-XSBi33OA7OPmF$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Jul 21 2026 - 08:01:49 MST