[AMMRL] Code for emailing when errors for Quantum Dynamics He recovery system? #HeliumRecovery

From: Emily Crull via groups.io <emily.crull=mail.wvu.edu_at_groups.io>
Date: Tue, 21 Jul 2026 09:35:57 -0700

Hi everyone,

We have a Quantum Dynamics helium recovery system and when it works, it is
an excellent system.

However, due to infrastructure issues, we often experience hotter-than-usable
water for the compressors. For reference, the system prefers <85F and we have
cooling issues that result in the water reaching 120F.

We're working on getting monitoring for the water that will email us if it
increases in temperature or decreases in pressure.

As a backup measure, I was wondering if anyone has written a script that monitors
the read outs of the liquifier and ATP and emails if the values for temperature
or pressure go outside certain acceptable ranges? For example, monitoring the
temperature in the liquifier and send an email if the temperature goes above 5K.

If you're willing to share the script, please let me know.

Many thanks,
Emily
Director of Chemistry Shared Instrumentation
C. Eugene Bennett Department of Chemistry
West Virginia University


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3602): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3602__;!!PvDODwlR4mBZyAb0!R2VZB9EQ2O5TyS0o-ITAW4K1AWiN6A=
wkh1EAXBdfWySaDwaRlbYuo54EJ0131jtwkMzpR-ukJ-jfv7jT2RUewFWjjH2P$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120380115=
/7559972__;!!PvDODwlR4mBZyAb0!R2VZB9EQ2O5TyS0o-ITAW4K1AWiN6Awkh1EAXBdfWySaD=
waRlbYuo54EJ0131jtwkMzpR-ukJ-jfv7jT2RUewDp7x_gg$
Mute #heliumrecovery:https://urldefense.com/v3/__https://ammrl.groups.io/g/=
main/mutehashtag/heliumrecovery__;!!PvDODwlR4mBZyAb0!R2VZB9EQ2O5TyS0o-ITAW4=
K1AWiN6Awkh1EAXBdfWySaDwaRlbYuo54EJ0131jtwkMzpR-ukJ-jfv7jT2RUewOrjs4Np$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 21 2026 - 09:42:52 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST