Re: [AMMRL] how to obtain a liquid 6Li NMR spectrum using Bruker NMR spectrometer

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Tue, 21 Jul 2026 14:37:25 +0000

Dear Junfeng,

the ability to run 6Li is depending on the hardware you have. 6Li is very
close to deuterium and if there is a 2H Stop filter in line you will not be
able to observe 6Li.

XBB19F or XBB preamplifiers come in two flavors with or without built in 2H
stop filter. They would have either HS or HP in their description.
It is also possible that the 2H stop is built into the probe. This is the case
for all BB cryoprobes and Prodigy probes. Some iProbes also have built in 2H
stop filters.

If there is no such filter in line you should be able to observe 6Li. If the
filter is external, remove it. If it is built in you are out of luck.

If you tell me which spectrometer you are using I can tell you if it will work.

With best regards

Clemens


On 7/20/2026 9:12 PM, 向俊锋 via groups.io wrote:

Dear colleagues

Is there anyone who has experience in using the Bruker liquid NMR spectrometer
to collect the 6Li spectrum and would like to share it?

Thank you very much!

Junfeng Xiang
NMR Lab
ICCAS, China




--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
ilto:clemens.anklin_at_bruker.com>
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3600): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3600__;!!PvDODwlR4mBZyAb0!Q0hx5J5g55p2ZiaDooouAFJ6kgH8dT=
ObBTBCqxet-SJF2paUd2KJVzhsCfroMH_25EVYP-in4dFzRC38R3M4VRrdjdra$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120376616=
/7559972__;!!PvDODwlR4mBZyAb0!Q0hx5J5g55p2ZiaDooouAFJ6kgH8dTObBTBCqxet-SJF2=
paUd2KJVzhsCfroMH_25EVYP-in4dFzRC38R3M4VeMWQ4UA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 21 2026 - 08:01:39 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST