Hi Jaya,
Here is some information. When you select connect to the spectrometer, the
software connects to the "linacq server". If the server wasn't shut down
properly you will get an error. So rebooting or running "linacq restart" in
a terminal will clear this.
Once you are connected properly you will see the normal display with the MW
bridge information and the field setting value.
Go from standby to operate. Is there a current value above zero? You can
move the bias slider to see if it changes.
If not, then move the frequency slider from one end to the other and see if
suddenly you now see diode current. If this happens then the Gunn source
oscillator is not oscillating at the initial set frequency and by changing
the frequency it will start.
Best regards,
Chuck
On Thu, Jul 16, 2026 at 7:31 PM Charles Hanson <hansoc_at_bc.edu> wrote:
> Hi Jaya,
> If you can’t get help from Bruker by Monday, I can help you with this. I
> will need to try to tune it manually to see what is going on.
> linacq restart” will restart the server so you don’t have to restart the
> computer. Also don’t connect until the blue light on the bridge stays
> solid.
> Best,
> Chuck
>
>
> Charles Hanson
> Assistant Director, Magnetic Resonance Center
> Boston College
> Merkert 107
> 2609 Beacon Street
> Chestnut Hill MA 02467-3860
> Tel (617) 552-6903 (office)
>
>
> https://www.bc.edu/content/bc-web/research/sites/vice-provost-for-research/about/research-facilities/nmr-center.html
>
>
> On Thu, Jul 16, 2026 at 7:24 AM Jaya Nolt via groups.io wrote:
>
>> Hi AAMRL group,
>>
>> Our EMXPlus Bruker EPR is having two issues (they may be related). I have
>> an email chain open with Bruker support but hoping for any advice in the
>> meantime.
>>
>> The first problem has been intermittent but is now persistent.
>> The frequency isn't locking when tuning or going to operate.
>> When auto tune is used or when I go from tune to standby the
>> frequency shifts from 9.8GHz to 12GHz. This throws Error 182 in the
>> bridge web page.
>>
>> The second problem I am now having is when I try to open Xenon it does
>> not initialize, throws errors about lack of communication. At first
>> rebooting the computer has helped. I have checked the ethernet cable from
>> the console to the computer. I have not replaced the hub.
>>
>> Some other details to note that may be useful,
>> 1) On start up the frequency is 0GHz, until I start the tuning process.
>> 2) The tuning dip does disappear after the 12GHz switch and I have to go
>> back to Standby first.
>> 3) We are running Xenon 1.2 on Centos 6.9, is it possible there may be
>> some firmware issues?
>>
>> thanks for any thoughts in advance,
>>
>> jaya
>>
>> --
>> Jaya Nolt
>> R & D Engineer
>> Spectroscopy Facility
>> Materials Research Lab
>> Office: (805) 893-4997
>> Email: jaya_at_mrl.ucsb.edu
>> http://www.mrl.ucsb.edu
>>
>>
>>
--
Charles Hanson
Assistant Director, Magnetic Resonance Center
Boston College
Merkert 107
2609 Beacon Street
Chestnut Hill MA 02467-3860
Tel (617) 552-6903 (office)
/urldefense.com/v3/__
https://www.bc.edu/content/bc-web/research/site=
s/vice-provost-for-research/about/research-facilities/nmr-center.html__;!!P=
vDODwlR4mBZyAb0!UPnumcvOWAbn-2J9EgE_czE86tCrYRRKXhYkzcPJhoUW1RyRJ16nyCtf-5P=
5mMsoKsV6ZtsDyX1eem2ZAodaThHx$
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3593): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3593__;!!PvDODwlR4mBZyAb0!UPnumcvOWAbn-2J9EgE_czE86tCrYR=
RKXhYkzcPJhoUW1RyRJ16nyCtf-5P5mMsoKsV6ZtsDyX1eem2ZAhyCVAd0$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120296777=
/7559972__;!!PvDODwlR4mBZyAb0!UPnumcvOWAbn-2J9EgE_czE86tCrYRRKXhYkzcPJhoUW1=
RyRJ16nyCtf-5P5mMsoKsV6ZtsDyX1eem2ZAtXcke7B$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Jul 20 2026 - 08:54:03 MST