Hi Jaya,
If you can’t get help from Bruker by Monday, I can help you with this. I
will need to try to tune it manually to see what is going on.
linacq restart” will restart the server so you don’t have to restart the
computer. Also don’t connect until the blue light on the bridge stays solid.
Best,
Chuck
Charles Hanson
Assistant Director, Magnetic Resonance Center
Boston College
Merkert 107
2609 Beacon Street
Chestnut Hill MA 02467-3860
Tel (617) 552-6903 (office)
https://www.bc.edu/content/bc-web/research/sites/vice-provost-for-research/about/research-facilities/nmr-center.html
On Thu, Jul 16, 2026 at 7:24 AM Jaya Nolt via groups.io wrote:
> Hi AAMRL group,
>
> Our EMXPlus Bruker EPR is having two issues (they may be related). I have
> an email chain open with Bruker support but hoping for any advice in the
> meantime.
>
> The first problem has been intermittent but is now persistent.
> The frequency isn't locking when tuning or going to operate.
> When auto tune is used or when I go from tune to standby the
> frequency shifts from 9.8GHz to 12GHz. This throws Error 182 in the
> bridge web page.
>
> The second problem I am now having is when I try to open Xenon it does not
> initialize, throws errors about lack of communication. At first rebooting
> the computer has helped. I have checked the ethernet cable from the console
> to the computer. I have not replaced the hub.
>
> Some other details to note that may be useful,
> 1) On start up the frequency is 0GHz, until I start the tuning process.
> 2) The tuning dip does disappear after the 12GHz switch and I have to go
> back to Standby first.
> 3) We are running Xenon 1.2 on Centos 6.9, is it possible there may be
> some firmware issues?
>
> thanks for any thoughts in advance,
>
> jaya
>
> --
> Jaya Nolt
> R & D Engineer
> Spectroscopy Facility
> Materials Research Lab
> Office: (805) 893-4997
> Email: jaya_at_mrl.ucsb.edu
> http://www.mrl.ucsb.edu
>
>
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3594): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3594__;!!PvDODwlR4mBZyAb0!VTZzvPR6lrLdr8_4V9Hi2grLqeBYyj=
Sc_jBqqWheWiKv5nitY5t9E3BuakNPnhdMul5A7_TbOI2YLJ3H_D4QsBzS$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120296777=
/7559972__;!!PvDODwlR4mBZyAb0!VTZzvPR6lrLdr8_4V9Hi2grLqeBYyjSc_jBqqWheWiKv5=
nitY5t9E3BuakNPnhdMul5A7_TbOI2YLJ3H_PJ9FZYY$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Jul 20 2026 - 08:54:29 MST