Hi AAMRL group,
Our EMXPlus Bruker EPR is having two issues (they may be related). I have
an email chain open with Bruker support but hoping for any advice in the
meantime.
The first problem has been intermittent but is now persistent.
The frequency isn't locking when tuning or going to operate.
When auto tune is used or when I go from tune to standby the
frequency shifts from 9.8GHz to 12GHz. This throws Error 182 in the bridge
web page.
The second problem I am now having is when I try to open Xenon it does not
initialize, throws errors about lack of communication. At first rebooting
the computer has helped. I have checked the ethernet cable from the console
to the computer. I have not replaced the hub.
Some other details to note that may be useful,
1) On start up the frequency is 0GHz, until I start the tuning process.
2) The tuning dip does disappear after the 12GHz switch and I have to go
back to Standby first.
3) We are running Xenon 1.2 on Centos 6.9, is it possible there may be some
firmware issues?
thanks for any thoughts in advance,
jaya
--
Jaya Nolt
R & D Engineer
Spectroscopy Facility
Materials Research Lab
Office: (805) 893-4997
Email: jaya_at_mrl.ucsb.edu
http://www.mrl.ucsb.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3574): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3574__;!!PvDODwlR4mBZyAb0!STXp8icS9XCBumIyHCSNQ677lng6Gg=
fsIFEy_3TROK1cgawPaUUiweRkhVJSjq98Q_38D49aSTjust40ycdg4Q3hrRHK$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120296777=
/7559972__;!!PvDODwlR4mBZyAb0!STXp8icS9XCBumIyHCSNQ677lng6GgfsIFEy_3TROK1cg=
awPaUUiweRkhVJSjq98Q_38D49aSTjust40ycdg4aAFv6x5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Jul 16 2026 - 04:24:34 MST