Re: [AMMRL] Lineshape anomaly

From: Shin, Ronald via groups.io <shinr=uab.edu_at_groups.io>
Date: Wed, 15 Jul 2026 21:22:03 +0000

We've had this issue with our 600 Mhz instrument, equipped with a SampleCase
and TCI Cryoprobe.

For the first 6 years, it worked great with the blue spinners, but I started
noticing that some of the samples weren't reaching the bottom all of the way,
since a previous deuterium signal that was very strong, looked very weak with
the next sample, despite sample conditions being the same (same solvent, same
brand and make of tube, similar samples).

I did find a work around, where I swapped out the blue 5mm spinners with the
white ones, which are a bit heavier, and those were properly dropping the
sample all the way into the bore.

Lately, though, sometimes the samples in the white spinners aren't reaching
the bottom of the bore, and get spit back up, dropped down, spat out, etc.
several times before either eventually it does settle into the bore, or it
just gets returned back to the SampleCase carousel.

Using the ceramic spinner works, but I figure it's probably time for a thorough
cleaning, since there's probably some gunk that's preventing the sample from
consistently seating in the bore.

Ronald Shin, PhD

University of Alabama at Birmingham
Department of Biochemistry and Molecular Genetics
Central Alabama High Field Nuclear Magnetic Resonance Facility




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3575): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3575__;!!PvDODwlR4mBZyAb0!QG2tZkXZz0DOFryH3dd5WCUwR0dmYa=
QZug9KyRfIKbWjOxofCAoMH3-T7JkN1KyjqAg3JMyizNwjqYDenvsVn2FF$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120287344=
/7559972__;!!PvDODwlR4mBZyAb0!QG2tZkXZz0DOFryH3dd5WCUwR0dmYaQZug9KyRfIKbWjO=
xofCAoMH3-T7JkN1KyjqAg3JMyizNwjqYDenhcXSkVe$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jul 16 2026 - 04:24:37 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST