Re: [AMMRL] Lineshape anomaly

From: 向俊锋 <jfxiang_at_iccas.ac.cn>
Date: Thu, 16 Jul 2026 11:39:37 +0800 (GMT+08:00)

Would you try to check the top of the shim coil as well as the probe? Somet=
imes,there were some dirty left after long time's running!


-----原始邮件-----
发件人:"Smith, Scott K via groups.io"
发送时间:2026-07-16 02:35:39 (星期四)
收件人: "main_at_ammrl.groups.io" <main_at_ammrl.groups.io> 抄送:
主题: [AMMRL] Lineshape anomaly


This is a strange but consistent problem that seems related to Bruker spinners
and just one of our automated spectrometers. Occationally a sample is inserted
and immediately shows ~1000Hz linewidth on what would otherwise be a normal
proton acquisition.


The samples are properly sized and set in the spinners.
The proton signal is strong, but lock is nearly gone.
Ejecting and reinserting has no effect.
The spinners themselves show no obvious defects.
Change the spinner and shimming goes back to normal like magic.


The problem was not reproducible on three other Bruker spectrometers with
and without automation.
A year ago this had become a near daily occurrence and we tried to isolate
the "bad" spinners. Eventually there were more bad than good spinners so
we ordered a new set for the sample case. For about six months the problem
was gone. Now it is starting back up at a rate of a failure every one or
two weeks. Now the "bad" spinners can be reused.

The console is an Avance III HD with a Prodigy probe ca 2018 and a sampleCase.
The magnet is a vintage unsheilded Oxford 500 ca 1982.
Has anyone encountered what I have?

keep 'em spinning
Scott


Scott K. Smith, Director of NMR Spectroscopy
The University of Texas at Austin | Chemistry | 832-294-1928 | utexas.edu
NMRlab homepage: IoNMR | Nuclear Magnetic Resonance Labs (utexas.edu)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3576): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3576__;!!PvDODwlR4mBZyAb0!RnFj6Q2-chrp7tQI-7bts5COaijaxU=
uFR1p_xhhJdZbUl4vBwEn2dUM0hpD3SLmPQoeiX9x8lkAdXcMcRw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120287344=
/7559972__;!!PvDODwlR4mBZyAb0!RnFj6Q2-chrp7tQI-7bts5COaijaxUuFR1p_xhhJdZbUl=
4vBwEn2dUM0hpD3SLmPQoeiX9x8lkB_d8_zvw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






1784173113681.png
(image/png attachment: 1784173113681.png)

1784173169240.png
(image/png attachment: 1784173169240.png)

Received on Thu Jul 16 2026 - 04:24:48 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST