From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Wed, 1 Jul 2026 15:18:18 +0000

Hi all,

I’m looking for some guidance on user permissions and expinstall behavior
in TopSpin.

I have a group who is operating under their own Linux account on the instrument.
When they run utilities like expinstall from their account, it appears to break
several of our automation scripts—likely due to permission or compilation issues.
To resolve this, we currently have to rerun expinstall under our NMRSU account
and ensure everything compiles properly before the user can continue working in
their own account.

I was advised that adding the user to an “NmrSuperUser” group would allow
them to perform these actions without causing issues. However, this group did
not exist on our system, so I created it and added both the user and
the NMRSU account. Unfortunately, this has not resolved the problem. Has anyone
dealt with this before or can clarify how expinstall permissions and compilation
are supposed to be handled across multiple user accounts?

Is there a recommended way to configure shared environments or permissions
so that users can safely run these utilities without breaking system-wide
functionality?

Thanks in advance for any insight.

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3529): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3529__;!!PvDODwlR4mBZyAb0!SMQIEbqljk5lFbkKheUf7LAa9tdoRf=
QRZw5txnUV0tqWpLgzzN2j926CMCWvkvstTfjvvZ9pFIx6hQUxBWxS2yUZ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120067317=
/7559972__;!!PvDODwlR4mBZyAb0!SMQIEbqljk5lFbkKheUf7LAa9tdoRfQRZw5txnUV0tqWp=
LgzzN2j926CMCWvkvstTfjvvZ9pFIx6hQUxBQclvGc6$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 01 2026 - 08:19:09 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST