Hi Tom,
you can use the AU program "sertoint" to convert 2D double data to integer.
Or you can set Topspin to do the conversion to integer at the end of the
acquisition by setting this option in setres.
Best regards
Clemens
On 7/24/2025 3:59 PM, Tom Pratum via groups.io wrote:
I am not sure if this is of value or not, but if you are trying to access the
data outside of TopSpin, the earlier TS data is 32 bit integer and the latter
is 64 bit float. We found this out when trying to read TS4.x data into an
older version of MNOVA, and had to upgrade that version to be successful.
Other methods of accessing the data outside of TS need to be similarly modified.
Tom Pratum, SOU
On Jul 24, 2025, at 7:39 AM, Delk, Michael Joseph via groups.io wrote:
More importantly, the Bruker rep told us there are issues or compatibility
problems with our older data from this magnet. This is old TS 2.6 data. Has
anyone found conversion methods or other ways to still make this older work
useful?
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2559): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2559__;!!PvDODwlR4mBZyAb0!VJGAVPRSeJTR9zA5zDjR20vzA3RFiT=
Rise7QA_otezzBf8F2M71omcfce7NYKk1Wd6fJS3nQXFi0LczIDp8wpfOTnZwr$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114322054=
/7559972__;!!PvDODwlR4mBZyAb0!VJGAVPRSeJTR9zA5zDjR20vzA3RFiTRise7QA_otezzBf=
8F2M71omcfce7NYKk1Wd6fJS3nQXFi0LczIDp8wpevyDq1y$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Jul 24 2025 - 14:26:54 MST