I am not sure if this is of value or not, but if you are trying to access the
data outside of TopSpin, the earlier TS data is 32 bit integer and the latter
is 64 bit float. We found this out when trying to read TS4.x data into an older
version of MNOVA, and had to upgrade that version to be successful. Other
methods of accessing the data outside of TS need to be similarly modified.


Tom Pratum, SOU


> On Jul 24, 2025, at 7:39 AM, Delk, Michael Joseph via groups.io wrote:
>
> More importantly, the Bruker rep told us there are issues or compatibility
> problems with our older data from this magnet. This is old TS 2.6 data.
> Has anyone found conversion methods or other ways to still make this older
> work useful?
>
>



-=-=-=-=-=-=-=-=-=-=-=-