Re: [AMMRL] a simple 2D TOCSY on a bruker cryoprobe

From: Shin, Ronald via groups.io <shinr=uab.edu_at_groups.io>
Date: Thu, 24 Jul 2025 14:51:31 +0000

Also,

I'd try running 'pulsecal' before running a "canned" experiment. This should
eliminate the possibility that you're exceeding the power levels, since all
of the other pulses (spinlock, etc) are calculated directly from the 90 degree
pulse in accordance to how your system was set up in the first place.

I just to see if you're still getting those warnings about power levels being
exceeded with a "canned" experiment (not a custom one), after 'pulsecal' has
been run.

Ronald Shin
University of Alabama at Birmingham
Central Alabama High Field NMR Facility
________________________________
> From: main_at_ammrl.groups.io on behalf of D'Anciaes Almeida Silva, Igor via groups.io
> Sent: Thursday, July 24, 2025 9:30 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] a simple 2D TOCSY on a bruker cryoprobe

Hi Serge,

I'd start checking the peak powers for the nuclei involved at "edprobe".
Maybe the ones at the pulse sequence are higher then what its written there.

Best,

----------------
Dr. Igor d'Anciães Almeida Silva
Research Scientist - NMR Core
University of Missouri-Columbia
32A Chemistry Building
________________________________
> From: main_at_ammrl.groups.io on behalf of serge_lavoie via groups.io
> Sent: Thursday, July 24, 2025 8:58 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] a simple 2D TOCSY on a bruker cryoprobe

Hi all,

I am trying to record a basic 2D TOCSY experiment with our instrument (Bruker
Neo 500 MHz, cryoprobe prodigy BBO). For a reason i dont understand, it fails
after a few dummy scan with this error message: 'EMERGENCY-STOP activated:
Peak power limit exceeded, EmStop activated...'. I have verified that the probe
was correctly tuned and matched. I have checked that all pulse length and power
are set according to the pulse program instruction. Here a the most important
parameter i used:


  * pulprog: dipsi2gpphzs
  * td = 4096 x 256
  * ds = 16, ns = 8
  * d1 = 2 s; d9 = 80 ms
  * sw = 2778 x 2778 Hz; o1p = 3.25 ppm
  * p1 = 10 µs, p6 = 30 µs, p32 = 20 ms
  * pldb1 = -11.92 db; pldb10 = -2.3807 db; spdb29 = 12.26 db

Note that using the same instrument, i successfully recorded 1H, 13C, hsqc,
hmbc, dqf-cosy, even hsqc-tocsy experiments. What is wrong with 2d tocsy?

Serge Lavoie
Université du Québec à Chicoutimi
Saguenay (Québec)




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2547): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2547__;!!PvDODwlR4mBZyAb0!R1tzZO4qkWG20ocJFjqXcLqz-dz2G-=
2sh36ulmHwIoQ5bIxWQjQ5f_HFqcvFyVWDpBuEBsLva1lKWUVuIacVf_Uh$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114321371=
/7559972__;!!PvDODwlR4mBZyAb0!R1tzZO4qkWG20ocJFjqXcLqz-dz2G-2sh36ulmHwIoQ5b=
IxWQjQ5f_HFqcvFyVWDpBuEBsLva1lKWUVuIS27-JD_$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jul 24 2025 - 10:44:12 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST