Hi all,
I am trying to record a basic 2D TOCSY experiment with our instrument (Bruker
Neo 500 MHz, cryoprobe prodigy BBO). For a reason i dont understand, it fails
after a few dummy scan with this error message: 'EMERGENCY-STOP activated: Peak
power limit exceeded, EmStop activated...'. I have verified that the probe was
correctly tuned and matched. I have checked that all pulse length and power
are set according to the pulse program instruction. Here a the most important
parameter i used:
* pulprog: dipsi2gpphzs
* td = 4096 x 256
* ds = 16, ns = 8
* d1 = 2 s; d9 = 80 ms
* sw = 2778 x 2778 Hz; o1p = 3.25 ppm
* p1 = 10 µs, p6 = 30 µs, p32 = 20 ms
* pldb1 = -11.92 db; pldb10 = -2.3807 db; spdb29 = 12.26 db
Note that using the same instrument, i successfully recorded 1H, 13C, hsqc,
hmbc, dqf-cosy, even hsqc-tocsy experiments. What is wrong with 2d tocsy?
*Serge Lavoie*