Re: [AMMRL] Bruker Avance II communication errors

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Wed, 17 Jul 2024 18:58:43 +0000 (UTC)

Hi Brian,

No, if the DHCP server's not running I don't think the NFS is going to make
much difference.  I would try starting with admin privileges.  If
it doesn't start you might look at your network configuration--if there's not
a connection called spect with IP address 149.236.99.1 and subnet mask
255.255.255.0 then DHCP won't start--it will look at the nework and say "I don't
recognize that" and shut down.  

Cheers,Mike

On Wednesday, July 17, 2024 at 12:11:07 PM MDT, Brian Antalek wrote:

Thx Michael,The Bruker DHCP service is stopped for some reason (see attached).
’Start’ is greyed out- do I restart here with admin privledges? 
I also ‘repaired' the open text NFS solo ver 14 app.  Didn’t
make a difference. 


On Jul 17, 2024, at 10:20 AM, Michael Groves via groups.io wrote:
 Hi Brian,
I'm not sure if grad is a useful tool on your instrument--the gradient amplifier
is likely in the BSMS so you might have to access it through the BSMS instead
of the grad tool you're using.  If you open the bsms tool and there's an
entry in the main menu for the gradients then you can check the gradient amp
through there.

Regarding the DHCP error, this usually means that one of the necessary processes
on the computer that talks to the console did not start when the computer powered
back up.  If this is a linux system, that stuff usually just works.  If
it's a windows system, have a look in the services that are running and check that
the Bruker DHCP and tftpd servers are running, and I think there's a process called
ccu that might need to be running to communicate with the CCU in the AQS rack.

Cheers,Mike

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1441): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1441__;!!PvDODwlR4mBZyAb0!TRyVA0ITVkqZLDDE9QtWwUD_BinLPL=
LrZGUnoUIXwATPE53-8ynb0I-IjTX3M_2cdbmcIUvIa6StzWf7VZ3LKhsu-xzo$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107271064=
/7559972__;!!PvDODwlR4mBZyAb0!TRyVA0ITVkqZLDDE9QtWwUD_BinLPLLrZGUnoUIXwATPE=
53-8ynb0I-IjTX3M_2cdbmcIUvIa6StzWf7VZ3LKrRw1sSI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 17 2024 - 12:01:53 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST