Re: [AMMRL] Bruker Avance II communication errors

From: Brian Antalek <bjantalek_at_gmail.com>
Date: Wed, 17 Jul 2024 13:59:50 -0400

Thx Michael,

The Bruker DHCP service is stopped for some reason (see attached). ’Start’
is greyed out- do I restart here with admin privledges?

I also ‘repaired' the open text NFS solo ver 14 app. Didn’t
make a difference.


> On Jul 17, 2024, at 10:20 AM, Michael Groves via groups.io wrote:
>
> Hi Brian,
>
> I'm not sure if grad is a useful tool on your instrument--the gradient amplifier
> is likely in the BSMS so you might have to access it through the BSMS instead
> of the grad tool you're using. If you open the bsms tool and there's an entry
> in the main menu for the gradients then you can check the gradient amp through there.
>
> Regarding the DHCP error, this usually means that one of the necessary processes
> on the computer that talks to the console did not start when the computer powered
> back up. If this is a linux system, that stuff usually justworks. If it's a windows
> system, have a look in the services that are running and check that the Bruker DHCP
> and tftpd servers are running, and I think there's a process called ccu that might
> need to be running to communicate with the CCU in the AQS rack.
>
> Cheers,
> Mike
>
> On Wednesday, July 17, 2024 at 07:50:03 AM MDT, Brian Antalek wrote:
>
>
> Hi all,
> I am having communication errors on my Bruker Avance II 300 MHz system (TopSpin 3.2).
> The problem occurred unassociated with any event (restart, powerdown, etc.).
>
> I have attached the error messages from 'ii' and 'ha' commands.
> Also, Below I have documented the feedback from the service tools, bsms.cmd and grad.cmd.
>
> Any ideas on what's wrong?
>
> Thanks!
> Brian Antalek
> Bausch+Lomb, Rochester NY
>
> ========================
> service tools
> bsms.cmd
> Init of BSMS ...
> is passing successfully!
> Press RETURN to continue...
>
> grad.cmd
> Init of Grad-CPU ...
> ERROR: G018 from GRT_: 10Corrupt data in BBIS 0 block 3, bus 1
> Wrong data in BBIS block 3. This block contains board
> specific data, like calibration values, parameters etc.
>
> Delete error? [y/n] y
> ERROR: G019 from GRT_: 00Corrupt data in BBIS 0 block 4, bus 0
> Wrong data in BBIS block 4. This block contains board
> specific data, like calibration values, parameters etc.
>
> Delete error? [y/n] y
> ERROR: G018 from GRT_: 00Corrupt data in BBIS 0 block 3, bus 0
> Wrong data in BBIS block 3. This block contains board
> specific data, like calibration values, parameters etc.
>
> Delete error? [y/n] y
> ERROR: 0039 from TOOL:
> function aborted after detected error in grad system
> Press RETURN to continue...




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1440): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1440__;!!PvDODwlR4mBZyAb0!QAfMa0updI3t5kn678vJ-6U17SEr0C=
Vb4oy5iUJQMYP6PJ-dPWuZOUYLyds5LX4mwUzKW4A7KUWrG5Re$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107271064=
/7559972__;!!PvDODwlR4mBZyAb0!QAfMa0updI3t5kn678vJ-6U17SEr0CVb4oy5iUJQMYP6P=
J-dPWuZOUYLyds5LX4mwUzKW4A7KXmqvepk$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 17 2024 - 11:11:04 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST