Re: [AMMRL] Agilent 7600-AS error -3128 encoder battery down

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 3 Jul 2024 17:33:05 +0200

Hi,

that is to help with the procedure to swap batteries: you plug in the new
battery into the empty plug and then only unplug the empty battery…

Cheers
Bert

> Am 03.07.2024 um 15:55 schrieb Holmes, Daniel <holmesd5_at_msu.edu>:
>
> One odd thing I found was that they zip tied two connectors together for
> the battery and bridged with the two with wires. Not sure why.
>
>
>> On Jul 3, 2024, at 9:43 AM, Spin-Doc wrote:
>>
>> Hi,
>>
>> I think it may be a good idea for you to check all connectors around the
>> controller board (unfortunately you need to remove the side plexiglass
>> panel as well as all cables/connectors in order to pull the board sled out) -
>> we had a good ten cables lose their isolation and making shorts all across.
>> Fixed them by carefully pulling out the connector pins out of their sockets
>> one by one (push down the tiny metal flap, then pull on the cable) and
>> re-isolate with shrink-tube.

>> I will send you a photo via PM how such cables may look like… (not
>> sure whether sending pictures to AMMRL is OK or not)
>>
>> Best regards
>> Bert
>>
>>> Am 03.07.2024 um 14:45 schrieb Holmes, Daniel
>>>
>>> Thanks, Bert, for the information and the offer. I found the CD hidden
> in the bottom of one of my lab drawers. I tested my old battery, and it was
> good as well. I am hoping that the problem is failure under load, which I
> haven’t tested. I spoke with Agilent tech support, and they mentiond
> that they had this error occur and simply going through the Cal_PP resolved
> it. Also, they sent me the procedure for replacing the battery. PM me if
> you would like to have it. Good luck on your repair.
>
> Dan
>>
>> Magnetic resonance equipment, service, training & more
>> Selling & buying used NMR equipment (all vendors)
>> Dr. rer. nat. Bert Heise
>> Owner, Applications Scientist
>>
>> Spin-Doc Bert Heise
>> Kiefernweg 13
>> 58239 Schwerte
>> Germany
>> Mobile: +49 171 5135315

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1392): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1392__;!!PvDODwlR4mBZyAb0!U7rDhT3Sysvp5HCZOCRBnjhH_haTW9=
MYVTDH_sWIaj1xpq0aAAFnxK1j4xnLL4_CiDZy3U5N2A1VXCQrOKts9A$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106921425=
/7559972__;!!PvDODwlR4mBZyAb0!U7rDhT3Sysvp5HCZOCRBnjhH_haTW9MYVTDH_sWIaj1xp=
q0aAAFnxK1j4xnLL4_CiDZy3U5N2A1VXCQkV9bMuw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Jul 03 2024 - 08:49:07 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST