One odd thing I found was that they zip tied two connectors together for the
battery and bridged with the two with wires. Not sure why.
On Jul 3, 2024, at 9:43 AM, Spin-Doc <bert.heise_at_spin-doc.net> wrote:
Hi,
I think it may be a good idea for you to check all connectors around the
controller board (unfortunately you need to remove the side plexiglass panel
as well as all cables/connectors in order to pull the board sled out) - we
had a good ten cables lose their isolation and making shorts all across.
Fixed them by carefully pulling out the connector pins out of their sockets
one by one (push down the tiny metal flap, then pull on the cable) and
re-isolate with shrink-tube.
I will send you a photo via PM how such cables may look like… (not
sure whether sending pictures to AMMRL is OK or not)
Best regards
Bert
Am 03.07.2024 um 14:45 schrieb Holmes, Daniel
Thanks, Bert, for the information and the offer. I found the CD hidden in the
bottom of one of my lab drawers. I tested my old battery, and it was good
as well. I am hoping that the problem is failure under load, which I haven’t
tested. I spoke with Agilent tech support, and they mentioned that they had this
error occur and simply going through the Cal_PP resolved it. Also, they sent
me the procedure for replacing the battery. PM me if you would like to have it.
Good luck on your repair.
Dan
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it)
MestreLab NMR Software (
http://www.mestrelab.com)
Barthel HF Amplifiers (
http://www.barthel-hf.de)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1391): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1391__;!!PvDODwlR4mBZyAb0!Qty5SJl0wPsEC3Rjna8KK-C97fzMF6=
mivNqxYgmSPG-j4tcLBlW2NngKMKQbOUlGtPwPadshyqTSaMle$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106921425=
/7559972__;!!PvDODwlR4mBZyAb0!Qty5SJl0wPsEC3Rjna8KK-C97fzMF6mivNqxYgmSPG-j4=
tcLBlW2NngKMKQbOUlGtPwPadshytwKggOz$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Jul 03 2024 - 08:28:18 MST