Re: [AMMRL] GAB_Error in Avance III BSMS #Bruker #Hardware

From: Ulrich <Ulrich.Haunz_at_uni-konstanz.de>
Date: Wed, 03 Jul 2024 12:17:01 +0200

Hello,

I had also troubles with the GAB board.

ha -> bsms -> gradients -> Gradient Amplifier Control

there on the control panel - are some buttons. There you can set the Gradient unit ON or OFF.
On the next page - you can do some diagnostic tests and Resets.

This helped me a lot. Or you can update the firmware for the GAB-Boad.
Via the BSMS Service Tools you can look, if the whole BSMS rack is OK or something is old.
Also you can easily access there the Firmware update routine for the GAB-Board, too.

gab2_fpga_00-013-02.bit is my actual firmware for the GAB/2 board.

The worst case is - the gradient is broken in the probe. There is a big round plug
outside of the probe.
This is the gradient plug. There you can easily measure the contacts through the
gradient. There should be a stable resistive value.
If it is broker - there are all contacts at 0 Ohm.

Yours sincerely,

Ulrich Haunz

Am Dienstag, Juli 02, 2024 21:02 CEST, schrieb "Bob Berno" <bberno_at_mcmaster=
.ca>:

> Hello Anne,
>
> I have seen this and similar errors a few times in the past.  On a few
> occasions it was a problem with the gradient circuit in the probe, and
> once it was due to a failure in the GAB board.
>
> From your symptoms, I think you have a problem with the GAB board.
> Do you know if the gradients work in any other experiments?
> What happens when you try to run an experiment with uses pulsed field
> gradients (PFGs) for coherence selection (most 2D experiments will be
> setup to use PFGs by default)? Do you see the lock level go down and
> then recover as the experiment runs?
>
> Good luck,
> Bob.
>
> On 2024-07-02 2:06 p.m., Anne Carroll via groups.io wrote:
> >
> > Hello,
> > Please let me know if you have encountered this issue and how your
> > resolved it.
> >
> > In TopSpin 3.6.2, the issue manifests this error "TopshimRawCheck -
> > the Z gradient does not work" whenever we run gradient shimming.
> > Non-gradient shimming (topshim tune) runs fine. There is nothing wrong
> > with the probe.
> > When I reboot the Gradient Amplifier Board GAB/2 from the Service
> > Page, I hear a click and the Ready light on the board turns green. If
> > I Refresh the web service page while the light is green, the Board
> > State claims 'Operation.' Within a few seconds, I hear another click
> > and the Ready light goes out. Now the Refresh shows 'GAB_Error.'
> >
> > Thank you,
> > Anne Carroll
> >
> > --
> > Anne M. Carroll, Ph.D. (she/her) NMR/MS Instrumentation Specialist
> > Department of Chemistry and Biochemistry Miami University Oxford, Ohio
> > USA513-529-3163 Office 678-637-6739 Cell carrol50_at_miamioh.edu
> >
>
> --
> Dr. Bob Berno
> Manager, NMR Facility
> McMaster University, ABB-156
> 1280 Main St. West
> Hamilton, ON L8S 4M1
> T: 905-525-9140 x24158 or x24686
> -------------------------------------------------------------------------
> This email contains CONFIDENTIAL information intended for the specific recipient(s).
> If you have an accommodation need for a planned meeting, please e-mail me directly and
> I will do my best to make appropriate arrangements.
> Should you require any materials sent via this e-mail address in an alternate/accessible
> format, please let me know.
>
> McMaster University sits on the traditional Territories of the Mississauga and
> Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.
>
>
>
>
--
=========================

Ulrich Haunz, Dipl.-Ing. (FH)
NMR-Spektroskopie

Raum L 710, L 807
NMR-Center M520, M521, M527
Postfach M 715
Fachbereich Chemie
Universitaet Konstanz

78457 Konstanz

Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
Tel.: +49 (0)170 9153213
FAX : +49 (0)7531 88-3898
=========================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1386): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1386__;!!PvDODwlR4mBZyAb0!Xo4mMOl43MRky7I-f1ohpCoJFcgrKB=
KxyErYkcrSto6Onmkoj54IUASSDnc-oE7k0OVf6K9OqJNsz5d2kjSvmdLKPqyHQQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107005267=
/7559972__;!!PvDODwlR4mBZyAb0!Xo4mMOl43MRky7I-f1ohpCoJFcgrKBKxyErYkcrSto6On=
mkoj54IUASSDnc-oE7k0OVf6K9OqJNsz5d2kjSvmdJO56CYhQ$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!Xo4mMOl43MRky7I-f1ohpCoJFcgrKBKxyErYkc=
rSto6Onmkoj54IUASSDnc-oE7k0OVf6K9OqJNsz5d2kjSvmdJASxxtow$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!Xo4mMOl43MRky7I-f1ohpCoJFcgrKBKxyE=
rYkcrSto6Onmkoj54IUASSDnc-oE7k0OVf6K9OqJNsz5d2kjSvmdJpITKZUQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Wed Jul 03 2024 - 03:17:15 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST