Re: [AMMRL] GAB_Error in Avance III BSMS #Bruker #Hardware

From: Bob Berno <bberno_at_mcmaster.ca>
Date: Tue, 2 Jul 2024 15:02:57 -0400

Hello Anne,

I have seen this and similar errors a few times in the past.  On a few
occasions it was a problem with the gradient circuit in the probe, and
once it was due to a failure in the GAB board.

From your symptoms, I think you have a problem with the GAB board.

Do you know if the gradients work in any other experiments?

What happens when you try to run an experiment with uses pulsed field
gradients (PFGs) for coherence selection (most 2D experiments will be
setup to use PFGs by default)? Do you see the lock level go down and
then recover as the experiment runs?

Good luck,
Bob.

On 2024-07-02 2:06 p.m., Anne Carroll via groups.io wrote:
>
>
> Hello,
> Please let me know if you have encountered this issue and how your
> resolved it.
>
> In TopSpin 3.6.2, the issue manifests this error "TopshimRawCheck -
> the Z gradient does not work" whenever we run gradient shimming.
> Non-gradient shimming (topshim tune) runs fine. There is nothing wrong
> with the probe.
> When I reboot the Gradient Amplifier Board GAB/2 from the Service
> Page, I hear a click and the Ready light on the board turns green. If
> I Refresh the web service page while the light is green, the Board
> State claims 'Operation.' Within a few seconds, I hear another click
> and the Ready light goes out. Now the Refresh shows 'GAB_Error.'
>
> Thank you,
> Anne Carroll
>
> --
> Anne M. Carroll, Ph.D. (she/her) NMR/MS Instrumentation Specialist
> Department of Chemistry and Biochemistry Miami University Oxford, Ohio
> USA513-529-3163 Office 678-637-6739 Cell carrol50_at_miamioh.edu
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1385): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1385__;!!PvDODwlR4mBZyAb0!R7210h8kjv7IcGeKzfbgPLpHmxupNl=
2f4bphNL-K4KAI8pav3hYSr1rG_FUgtnrjduItNvpeuJKHMalH$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107005267=
/7559972__;!!PvDODwlR4mBZyAb0!R7210h8kjv7IcGeKzfbgPLpHmxupNl2f4bphNL-K4KAI8=
pav3hYSr1rG_FUgtnrjduItNvpeuCKIXhNM$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!R7210h8kjv7IcGeKzfbgPLpHmxupNl2f4bphNL=
-K4KAI8pav3hYSr1rG_FUgtnrjduItNvpeuBBW4TQc$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!R7210h8kjv7IcGeKzfbgPLpHmxupNl2f4b=
phNL-K4KAI8pav3hYSr1rG_FUgtnrjduItNvpeuEad_UCy$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 02 2024 - 12:03:07 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST