Re: [AMMRL] GAB_Error in Avance III BSMS #Bruker #Hardware

From: Parnham, Stu <stitch_at_email.unc.edu>
Date: Tue, 2 Jul 2024 18:28:34 +0000

Hey Anne,

I find rebooting the ELCB board via 'ha' clears the problem.

Stu

-------------------
Stuart Parnham Ph.D.
Biomolecular NMR
Marsico Hall, SB205A
University of North Carolina
Phone:(919)-966-6184
> From: main_at_ammrl.groups.io On Behalf Of Anne Carroll via groups.io
> Sent: Tuesday, July 2, 2024 2:06 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] GAB_Error in Avance III BSMS #Bruker #Hardware

Hello,
Please let me know if you have encountered this issue and how your resolved
it.

In TopSpin 3.6.2, the issue manifests this error "TopshimRawCheck - the Z
gradient does not work" whenever we run gradient shimming. Non-gradient
shimming (topshim tune) runs fine. There is nothing wrong with the probe.
When I reboot the Gradient Amplifier Board GAB/2 from the Service Page, I
hear a click and the Ready light on the board turns green. If I Refresh the =
web service page while the light is green, the Board State claims 'Operation.'
Within a few seconds, I hear another click and the Ready light goes out.
Now the Refresh shows 'GAB_Error.'

Thank you,
Anne Carroll

--
Anne M. Carroll, Ph.D. (she/her) NMR/MS Instrumentation Specialist
Department of Chemistry and Biochemistry
Miami University
Oxford, Ohio USA
513-529-3163 Office
678-637-6739 Cell
carrol50_at_miamioh.edu<mailto:carrol50_at_miamioh.edu>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1384): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1384__;!!PvDODwlR4mBZyAb0!RO49jjjyuxIwLgNQHcszXzyFdjPNib=
65EkrlSol6keX6cgPWwU6DXnUqr9a3U9v_tiwcTJqcIs5rdxMo8pw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107005267=
/7559972__;!!PvDODwlR4mBZyAb0!RO49jjjyuxIwLgNQHcszXzyFdjPNib65EkrlSol6keX6c=
gPWwU6DXnUqr9a3U9v_tiwcTJqcIs5rWxj-Rrc$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!RO49jjjyuxIwLgNQHcszXzyFdjPNib65EkrlSo=
l6keX6cgPWwU6DXnUqr9a3U9v_tiwcTJqcIs5rQcEPZV0$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!RO49jjjyuxIwLgNQHcszXzyFdjPNib65Ek=
rlSol6keX6cgPWwU6DXnUqr9a3U9v_tiwcTJqcIs5rs0WUZRs$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 02 2024 - 11:28:45 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST