Thank you all for your valuable information.
Since AlmaLinux 9 has some compatibility issues with the software I need to
run, I installed CentOS 7 on an HP Z420 workstation. With that older system,
both TopSpin 3.6 and IBS run normally.
Therefore, I will keep the AlmaLinux 9 computer as a backup system for all
NEO consoles.
The package tk.x86_64 was already installed, but installing TopSpin 4.4.1 gives
the following error:
NOTE: initial log file: /tmp/install.log
*** Fatal error ***
can't find package Tk
Details:
while executing
"package require Tk"
(procedure "ScanArguments" line 42)
invoked from within
"ScanArguments"
(procedure "MainProgram" line 9)
invoked from within
"MainProgram"
Any help or suggestions would be greatly appreciated.
Best regards,
Weixing Zhang
> >From: main_at_ammrl.groups.io On Behalf Of Anklin, Clemens via groups.io
> Sent: Wednesday, January 21, 2026 4:09 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] AlmaLinux 9, TopSpin 3.6/3.7, and Desktop Icons Questions
Hi Weixing
we have not validated TOpspin 3.6 with Alma 9. We did do this with Topspin
3.8. This is the current version for AV III and HD spectrometers
We do recommend XCFE as desktop environments
Sorry I can not help with the IBS question.
Regards
Clemens
On 1/21/2026 3:27 PM, Walter Zhang via groups.io wrote:
Dear colleagues,
We are planning to upgrade the computer for an AV III HD system,
and I would like to ask for your experience with running TopSpin 3.6
on AlmaLinux 9
In particular, I would greatly appreciate your feedback on the following questions:
1. TopSpin 3.6 on AlmaLinux 9
What has been your experience running TopSpin 3.6 on AlmaLinux 9?
Have you encountered any issues related to stability, performance, or software compatibility?
2. Desktop icons in AlmaLinux 9
The default Desktop environment (Standard, X11 display server) in AlmaLinux 9
as configured by Bruker does not display desktop icons.
Has anyone successfully used XFCE or KDE as the desktop environment for operating
the NMR spectrometer?
3. IBS compatibility with TopSpin 3.7
I previously attempted to use IBS with TopSpin 3.7 but was unable to
get it working.
Has anyone successfully configured IBS with TopSpin 3.7, and if so,
are there any special configuration steps or recommendations?
Any shared experiences, insights, or recommendations would be most appreciated.
Thank you in advance for your help.
Best regards,
Weixing Zhang
St. Jude Children’s Research Hospital
Memphis, TN 38105
USA
-- =========================
==========================
=====================
Clemens Anklin Ph.D. Vice President Bruker BioSpin
NMR Applications & Training 15 Fortune Drive
NEW Phone: 978-313-5NMR(5667) Billerica MA 01821
bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========
==========================
==========================
===========
________________________________
Email Disclaimer: http://www.stjude.org/emaildisclaimer
Consultation Disclaimer: http://www.stjude.org/consultationdisclaimer
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3157): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3157__;!!PvDODwlR4mBZyAb0!ShBReBD3tz47tpw77vdxscgz73iJTF=
1CWSiASCfh8GSH78m473Y2dA5061UL1xqpfwyqpq8z-UlEKnXDn6qK7vkKqwdH$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117388545=
/7559972__;!!PvDODwlR4mBZyAb0!ShBReBD3tz47tpw77vdxscgz73iJTF1CWSiASCfh8GSH7=
8m473Y2dA5061UL1xqpfwyqpq8z-UlEKnXDn6qK7ugZ10FB$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Jan 27 2026 - 03:35:53 MST