I can confirm these …
1) the error with the HPPR keeps coming back
2) I cannot telnet to spec or ccu
3) there is no indication of power failure at the rear of the AQS
4) I tried the red reset button on the ccu with no effect
5) I can connect to the DRU through a browser, but I couldn’t find
anything interesting there.
I will look into Microsoft breaking something on the firewall, but I think
I turned the firewall off.
Thanks for the attention here
-Brian
> On Jan 14, 2026, at 3:11 PM, Anklin, Clemens via groups.io wrote:
>
> Hi Brian
>
> could a recent update on the workstation have touched the firewalls? If the
> firewall rules for the spectrometer port got modified this could cause
> communication issues.
>
> Clemens
>
> On 1/14/2026 2:25 PM, Brian Antalek via groups.io wrote:
>> I have a Bruker AVII 300MHz instrument that is having communication problems.
>> 1) cf hangs on iiconf:connect to CCU and there is a message that reads,
>> connection to spec (aqport0) failed unable to connect to remote host
>> 2) the HPPR/2 unit has an error, error 13 power fail
>> 3) both SGU boards have a blinking error light
>> 4) the service tools report communication errors
>> 5) He level meter, lock, etc. - all not working
>>
>> I replaced the ethernet switch with another one that is known to be good
>> but no luck, same errors.
>>
>> any ideas?
>>
>> thanks
>>
>> -Brian
>>
>> Brian Antalek
>> Bausch + Lomb, Inc.
>> Rochester, NY
>> brian.antalek_at_bausch.com <mailto:brian.antalek_at_bausch.com>
>>
>
> --
========================
==========================
======================
Clemens Anklin Ph.D. Vice President Bruker
BioSpin NMR Applications & Training 15 Fortune Drive
NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
=========================
=========================
====================
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3125): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3125__;!!PvDODwlR4mBZyAb0!VFaSxR1PdTuveXZzFjiMRUk5SubhjG=
QOtkHjqQ9sDprJ-Zsoy4JGuc_KU_efX_398GCXhr6mRFX9J-bJNgeVfW7bXkw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117267180=
/7559972__;!!PvDODwlR4mBZyAb0!VFaSxR1PdTuveXZzFjiMRUk5SubhjGQOtkHjqQ9sDprJ-=
Zsoy4JGuc_KU_efX_398GCXhr6mRFX9J-bJNgeVKpqPWbI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Jan 14 2026 - 18:04:13 MST