Re: [AMMRL] Emergency lift won't stop (AvIII)

From: Todd Rappe via groups.io <rapp0006=umn.edu_at_groups.io>
Date: Mon, 5 Jan 2026 11:47:21 -0600

This post reminded me of a problem I had last summer. My symptoms were
vacuum errors and I troubleshooted (troubleshot?) it to the V11 solenoid
control valve inside the cryoplatform. Bruker quoted me for a replacement
of the whole pneumatic assembly (pictured below) but I ended up getting
just the individual control valve from Grainger and replacing the V11/V12
solenoid control valve. V13 is the emergency lift, and since the V15 valve
seems to impact the performance, I would be suspicious of the V13/V15
solenoid control valve.


_____________________________________
Todd Rappe
Minnesota NMR Center <nmr.umn.edu>
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3095): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3095__;!!PvDODwlR4mBZyAb0!WL9M6VdZDN4IgTOOwRqXX_SXoxhxEU=
4AKUbQTk88MgpbhkczCs40ji4q9-cKNZDRDPagk-pvMA1QEzc2kpZISgpyNt-I$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116735301=
/7559972__;!!PvDODwlR4mBZyAb0!WL9M6VdZDN4IgTOOwRqXX_SXoxhxEU4AKUbQTk88Mgpbh=
kczCs40ji4q9-cKNZDRDPagk-pvMA1QEzc2kpZISrTdyHne$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jan 05 2026 - 10:21:43 MST

This archive was generated by hypermail 2.4.0 : Mon Feb 09 2026 - 16:54:28 MST