Hi Lenny,
I’ve seen this error message on our system when the sample “catcher”
prongs fail to retract after a sample is ejected. This can happen because
there is too much friction for the return spring in the solenoid to overcome,
which in turn on our system originated from an alignment problem with the
carousel rack.
When you power up the unit it will do a self-test where you can (hopefully)
see its hesitance to retract back into the unit. If you apply pressure to
the lower corners of the rack relative the rest of the unit, one side should
be able to magically make it unstick.
On our system I got our mechanical workshop to progressively shave off small
amounts of material off of the offending plastic feet on the base of the rack
(sorry no pictures), which gives the system a little bit more wiggle room when
the locating elements on the back of the sample changer engage on startup.
We were having issues with our sample changer so the back part got replaced, and
apparently that alignment between older and newer units is not quite the same,
and the tolerances on the feet aren’t quite forgiving enough to handle it.
Haven’t seen this problem in ages so I can confirm it does the trick.
Thanks
Geoff
> From: main_at_ammrl.groups.io On Behalf Of Lenny Haddad via groups.io
> Sent: 20 February 2026 09:50
> To: main_at_ammrl.groups.io
> Subject: [External] [AMMRL] Frequent SampleXpress transfer failures AVANCE NEO 400 MHz
Hi,
We are experiencing frequent SampleXpress transfer failures on a 400 MHz
AVANCE NEO spectrometer running TopSpin 4.4.0 in continuous multi-user
automation.
The reported error is:
“Sample release could not be activated,” with the sample release state
shown as “Undefined” or “Inactive”.
No tube is physically stuck in the magnet when this occurs, and the issue is
not associated with a specific sample or user. Restarting the automation
system temporarily resolves the problem.
Do you have any idea why we might be experiencing these failures?
Thank you for your help.
Best regards,
Lenny
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3249): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3249__;!!PvDODwlR4mBZyAb0!T6RUYkL5GEvUYlclvkgCRzshgf7PHe=
Fe3DidOJtRYdIspJzG0-CDLOr3hJsVJh9MYRuBDwU4zei0oh6AOQMMow0lI1yT$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117909328=
/7559972__;!!PvDODwlR4mBZyAb0!T6RUYkL5GEvUYlclvkgCRzshgf7PHeFe3DidOJtRYdIsp=
JzG0-CDLOr3hJsVJh9MYRuBDwU4zei0oh6AOQMMoxw4ZCYn$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Feb 20 2026 - 06:15:53 MST