Hi Lenny,
This error pops up when the pin actuator either doesn’t fully come
forward into the hole in the cassette to release the sample, or if the
microswitch that is depressed when the actuator moves forward is failing. If
I were you, I’d check whether the silver pin is aligning with the
hole in the cassette and also try removing the cassette and switching the
sample release in service mode on the sample xpress. There is also a spring
in this mechanism which can break. We have replaced this spring in the past
and recovered from a similar error. I am happy to discuss more if you’d like.
Best of luck,
Sarah
Sarah Willis
Operations Manager
Department of Chemistry Instrumentation Facility (DCIF)
Massachusetts Institute of Technology
21 Ames St. Rm: 18-0084
Cambridge, MA 02142
sarahjbw_at_mit.edu<mailto:sarahjbw_at_mit.edu>
> From: main_at_ammrl.groups.io on behalf of Lenny Haddad via groups.io
> Date: Friday, February 20, 2026 at 8:26 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Frequent SampleXpress transfer failures – AVANCE NEO 400 MHz
Hi,
We are experiencing frequent SampleXpress transfer failures on a 400 MHz
AVANCE NEO spectrometer running TopSpin 4.4.0 in continuous multi-user
automation.
The reported error is:
“Sample release could not be activated,” with the sample
release state shown as “Undefined” or “Inactive”.
No tube is physically stuck in the magnet when this occurs, and the issue is
not associated with a specific sample or user. Restarting the automation
system temporarily resolves the problem.
Do you have any idea why we might be experiencing these failures?
Thank you for your help.
Best regards,
Lenny
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3248): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3248__;!!PvDODwlR4mBZyAb0!QJtM7oZk52_6OzErS11m3WF0ItTjcV=
ZILJB8toxCZCQDa8UaSQgskbSBhj1UFOzuI1p09xT-i-ckRTs7RVdxyL_4TmRb$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117909328=
/7559972__;!!PvDODwlR4mBZyAb0!QJtM7oZk52_6OzErS11m3WF0ItTjcVZILJB8toxCZCQDa=
8UaSQgskbSBhj1UFOzuI1p09xT-i-ckRTs7RVdxyIlVEAhM$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Feb 20 2026 - 06:05:22 MST