Hi,
We have TopSpin 4.4 running in Alma Linux 9.x and we have also noticed
the same kind of problem that you described, where IconNMR freezes and
doesn't allow to change anything in the queue. You can login and logout
of Icon, but nothing else. So I don't think this is related to your
broken Windows driver. In linux, I use xkill to kill IconNMR, but I
have been wandering if there is a solution or a better way to deal with
this problem. I would also like to know what is the cause of this problem.
Eugenio
--
Eugenio Alvarado, Ph.D.
NMR Facility Director
Chemistry Department, room 3500. University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924
On 2/18/26 2:48 PM, T. Andrew Mobley via groups.io wrote:
> Hi all,
>
> We have had a recent Windows 10 —> 11 update that has broken a driver.
> A quick google indicates that this is an old issue that we have just
> discovered. The driver is the iqvw64e.sys and it appears to have
> something to do with networking.
>
> I am just wondering if anyone knows if this will affect
> Topspin/IconNMR. We have recently had some weird freezes of IconNMR
> where the queue is still running but the interface won’t let us enter
> in any new experiments and does not respond to clicks either. Our
> solution has been to exit Topspin and restart IconNMR. We usually let
> the queue finish off its work before doing so.
>
> I was wondering if anybody had insight as to whether this driver could
> be the problem, and if there is a workaround if it is indeed the
> problem. This is the sole “new” issue that we can point a finger at so
> far.
>
> Another possibility is that a user occasionally puts an entry that
> contains a space in the “Name” field. We know that this messes with
> some archiving operations, but wondered if this could be part of the
> problem. This is HARDLY a new issue, however, so we don’t think it is
> likely the root of our freezing problem.
>
> I am currently in the process of scouring logs to see if I can find
> some pattern, but I thought I would ask the collective mind if someone
> has an idea.
>
> Best,
>
> Andy
>
>
> T. Andrew Mobley
>
> Personal Pronouns: He/Him/His
>
>
>
> Professor of Chemistry
> Noyce Science Center 2606
> 1116 8^th Avenue
> Grinnell College
> Grinnell, IA 50112
> 641-269-4543
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3242): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3242__;!!PvDODwlR4mBZyAb0!X2_Tak_IsTPdjMBsWCj8dnJ6u6vUBn=
LLb6IM84SrJYjlCbsfNLMo0zYNTo1gxap5NYnODnASa5X-KiYCaIfuXDBFXlQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117881085=
/7559972__;!!PvDODwlR4mBZyAb0!X2_Tak_IsTPdjMBsWCj8dnJ6u6vUBnLLb6IM84SrJYjlC=
bsfNLMo0zYNTo1gxap5NYnODnASa5X-KiYCaIfuv5K3Guw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
From owner-ammrl_at_ammrl.groups.io WedThu, 19 Feb 2026 03:51:25
Received: by 2002:a05:7508:526:b0:dc:b256:7258 with SMTP id j38csp821662gbb;
Thu, 19 Feb 2026 03:51:25 -0800 (PST)
X-Forwarded-Encrypted: i=2; AJvYcCVwXzJEF3bKGYVRJJwGDWPTKl5Q4CRLwgU+5kpz+8LxUSeTtdg5DwluusJilQQVLliHU9si_at_hawaii.edu
X-Received: by 2002:a05:690e:4091:b0:649:45c2:6d62 with SMTP id 956f58d0204a3-64c6136ca08mr1929136d50.5.1771501885249;
Thu, 19 Feb 2026 03:51:25 -0800 (PST)
ARC-Seal: i=1; a=rsa-sha256; t=1771501885; cv=none;
d=google.com; s=arc-20240605;
b=br7lKB+wH31dH2QDY9+Kkt0DUMq1Ro+Ll+mHQEisr3bhkEU6+4NE2/iKEsLFFKpTTv
sErKU16ayaERIgTh5IQ4mtdM07AarvDGdDRlowjB/InXJPl/Zi/z0xpGfVdrgv6e0pfH
+iEBrdSNJhs9fQqR0HMQXQf1jlsAOw249+E7ForQBBQhZlHPH2FmNlzX86xb0A7krIp6
+l1HB5URc0vud2JM8HgADjoDFC7j/NBRIjr1gxA/bjazGzT16xhUKBI4noDCeK5IZJEE
GQMeO3/u7m2RnXTe5qrNX2gY1Dn8blkTQPzjyqTt7hs8W5CMj7R+RgrdQvPzcWeLpuSk
gG2w==
ARC-Message-Signature: i=1; a=rsa-sha256; c=relaxed/relaxed; d=google.com; s=arc-20240605;
h=mime-version:content-language:list-unsubscribe
:list-unsubscribe-post:reply-to:resent-from:resent-date:delivered-to
:mailing-list:list-id:sender:list-help:list-subscribe:precedence
:accept-language:in-reply-to:references:message-id:date:thread-index
:thread-topic:subject:to:from:dkim-signature:resent-message-id;
bh=/EwMUcb8S3bodS7taVA9omsP1yl0Emp+wwf+oju/93g=;
fh=rQCyWLtLf8eHlGWzYFK0FoTKvDIIUhZg0EJcbtgap+A=;
b=OVTopzzR74/k2HC5hK5XSwlhXzwqUg+mC9fSEYOzw3WYHDkgiRBJaZM2AYOv4HR74X
6y76imL1LNkzqvqIPjPPJMj8gGpwydvGUX7cDb5sVzuQsKZ/HAlZFOip7Wn98cSnfIWY
//RwRxrB2J/aMXge37LSdxNmbjaie4dmjhEbmvzq4Qnrxc4ir9xvrp7OzC36C41gzMue
4T77l/Kgn6VXX7tqWDIurqHckjH1v8jAD1Qk6qO+3q/IuB4JcwTJIcj4eTNM730VU4AQ
ksVYfozZuRyKsMShI3W+gGG7Jv2mMlN3Sh2b0N3vWavXzSdr4R0cfOpnoU11kF5uI3TK
PFSg==;
dara=google.com
ARC-Authentication-Results: i=1; mx.google.com;
dkim=neutral (body hash did not verify) header.i=_at_groups.io header.s=groups1 header.b=Su1Ew1or;
spf=pass (google.com: domain of bounce+115296+3245+7559972+12177410_at_groups.io designates 192.53.124.254 as permitted sender) smtp.mailfrom=bounce+115296+3245+7559972+12177410_at_groups.io
Return-Path: <bounce+115296+3245+7559972+12177410_at_groups.io>
Received: from mx0b-007a6701.pphosted.com (mx0b-007a6701.pphosted.com. [205.220.181.169])
by mx.google.com with ESMTPS id 956f58d0204a3-64c22e07b74si11180033d50.11.2026.02.19.03.51.25
for <walt_at_hawaii.edu>
(version=TLS1_3 cipher=TLS_AES_256_GCM_SHA384 bits=256/256);
Thu, 19 Feb 2026 03:51:25 -0800 (PST)
Received-SPF: pass (google.com: domain of bounce+115296+3245+7559972+12177410_at_groups.io designates 192.53.124.254 as permitted sender) client-ip=192.53.124.254;
Authentication-Results: mx.google.com;
dkim=neutral (body hash did not verify) header.i=_at_groups.io header.s=groups1 header.b=Su1Ew1or;
spf=pass (google.com: domain of bounce+115296+3245+7559972+12177410_at_groups.io designates 192.53.124.254 as permitted sender) smtp.mailfrom=bounce+115296+3245+7559972+12177410_at_groups.io
Received: from pps.filterd (m0313028.ppops.net [127.0.0.1])
by mx0b-007a6701.pphosted.com (8.18.1.11/8.18.1.11) with ESMTP id 61J9tTSV2516443
for <walt_at_hawaii.edu>; Thu, 19 Feb 2026 01:51:25 -1000
Resent-Message-Id: <202602191151.61J9tTSV2516443_at_mx0b-007a6701.pphosted.com>
Authentication-Results: ppops.net;
spf=pass smtp.mailfrom=bounce+115296+3245+7559972+12177410_at_groups.io;
dkim=pass header.d=groups.io header.s=groups1;
dmarc=pass header.from=groups.io
Received: from mail08.groups.io (mail08.groups.io [192.53.124.254])
by mx0b-007a6701.pphosted.com (PPS) with ESMTPS id 4cdv3kgwxb-1
(version=TLSv1.2 cipher=ECDHE-RSA-AES128-GCM-SHA256 bits=128 verify=NOT)
for <walt_at_hawaii.edu>; Thu, 19 Feb 2026 01:51:24 -1000 (HST)
DKIM-Signature: a=rsa-sha256; bh=oObjiLSnkBeycpNvpGr3eOm8AvaxwncsVGmfUCIo4Wo=;
c=relaxed/relaxed; d=groups.io;
h=From:Subject:Date:Message-ID:To:Cc:In-Reply-To:References:Reply-To:Sender:List-Id:List-Post:List-Subscribe:List-Unsubscribe:List-Help:MIME-Version:Content-Type:Content-Transfer-Encoding;
s=groups1; t=1771501882; v=1; x=1771761082;
b=Su1Ew1orSrv7eze5VgBkVTYtkc9c7vPjOzG77KZxmAzll3JFNVsZEIKSpeHO9i9rc+yaHD7T
E6QZaAWQlBIGxrtaTwpB+R5GnK72eF+RxPL6MpIS3iw9dV3+kQl476x/rQP48+ry5fZMEzMfQ1B
//a6G0muAvolxJWhSxXNs/NDJ74FQ+zxUfFXCejALxzBy1Z8A/hAd43coeAjaWhXyfomAkiN63e
e8dl5AUj31440HfVlYndIFpoV1o/pfoAh2NACLnE/bL9uP/7GrVAso9gwUySu9VO6LHscPAXrIH
dPHocJSRc0s7oEN3N+5o39AGMwStsN3q7rTKIcXmV/MLQ==
X-Received: by 127.0.0.2 with SMTP id QXSCYY7561307xpqBX3XJwvE; Thu, 19 Feb 2026 03:51:22 -0800
X-Received: from DUZPR83CU001.outbound.protection.outlook.com (DUZPR83CU001.outbound.protection.outlook.com [52.101.66.120])
by mx.groups.io with SMTP id smtpd.msgproc02-g2.10423.1771494503718105216
for <main_at_ammrl.groups.io>;
Thu, 19 Feb 2026 01:48:25 -0800
X-Received: from AM6PR01MB5384.eurprd01.prod.exchangelabs.com
(2603:10a6:20b:9f::22) by GV1PR01MB10739.eurprd01.prod.exchangelabs.com
(2603:10a6:150:1ee::11) with Microsoft SMTP Server (version=TLS1_2,
cipher=TLS_ECDHE_RSA_WITH_AES_256_GCM_SHA384) id 15.20.9632.14; Thu, 19 Feb
2026 09:48:17 +0000
X-Received: from AM6PR01MB5384.eurprd01.prod.exchangelabs.com
([fe80::54e6:8065:24d8:42fe]) by AM6PR01MB5384.eurprd01.prod.exchangelabs.com
([fe80::54e6:8065:24d8:42fe%4]) with mapi id 15.20.9632.010; Thu, 19 Feb 2026
09:48:17 +0000
From: "Gary Thompson via groups.io" <G.S.Thompson=kent.ac.uk_at_groups.io>
To: "main_at_ammrl.groups.io" <main_at_ammrl.groups.io>
Subject: Re: [AMMRL] OCXO REF1200 board
Thread-Topic: [AMMRL] OCXO REF1200 board
Thread-Index: AQHcoKhPtfvzPhsQLUCdiSre7Ho13bWImLeAgAAY9wCAABDsAIABBccA
Date: Thu, 19 Feb 2026 09:48:17 +0000
Message-ID: <447997EC-3020-490D-A99F-A8622A1796CA_at_kent.ac.uk>
References: <D5DF2D2F-38E7-4B78-AE08-5DD637248B9C_at_uibk.ac.at>
<89f08ca0-0b89-4ecf-a3df-1967d4b7d18b_at_mcmaster.ca>
<764F8460-76FB-4794-A469-216F6A10E4D2_at_kent.ac.uk>
<F51AF820-B457-40C7-984A-CA8438AFB083_at_uibk.ac.at>
In-Reply-To: <F51AF820-B457-40C7-984A-CA8438AFB083_at_uibk.ac.at>
Accept-Language: en-GB, en-US
X-MS-Has-Attach: yes
X-MS-TNEF-Correlator:
x-ms-publictraffictype: Email
x-ms-traffictypediagnostic: AM6PR01MB5384:EE_|GV1PR01MB10739:EE_
x-ms-office365-filtering-correlation-id: fc0db64e-eba3-46b9-99ff-08de6f9c0236
x-ms-exchange-atpmessageproperties: SA
x-ms-exchange-senderadcheck: 1
x-ms-exchange-antispam-relay: 0
x-microsoft-antispam-message-info:
=?utf-8?B?bXBWWURhLy9xdDlFNEhTYTN4SmtSRHBaTXc5bmUrZi9JWDJwN1FFRk5Xcktw?=
=?utf-8?B?cGFpNExmSWpqaWE4cm9lbUVTTHdkbkFRNjE4MU0raWlvTzJKenI4ZTYyenF1?=
=?utf-8?B?azI1THJUaEtEb21YdnNGQis1eXEwL0hYb1kyK25PT1owTGZqSGJiSkNkK2p4?=
=?utf-8?B?eW1jMFJiY3hJam9CdFBDRXNGS0pJckVnMkk0V3l0NmZ0SUIxWmFicW1sM2RD?=
=?utf-8?B?cWsyd0JZank3NjYwcUR4N0twb2hRdlBPRUFvTkJ5a3ZzVGRwTnY3Y1JqZjcz?=
=?utf-8?B?bG4vTk53VUpHbHpFNHRjQmhIa1ZvVkRVSnp5dTlmNXNYclM4cVFiVTF6a2ow?=
=?utf-8?B?L2R0OXFCS01GdmZLY0l6QVdFajNDaTFDWE9pMjVPVlg2WGRmWTR1WTI0R2hw?=
=?utf-8?B?clJRTmNiZ01PY1pHa2tQUURTd3NtS3d5YW1HZFlQRkw4VHJFKzROem1ZTVZz?=
=?utf-8?B?ZmVJdElXaEMya2VUNUkyS3NYMmVIMVJSL0pkQU5wS3lDMlh4NFNUb1VsbE0x?=
=?utf-8?B?WEZFZHUwNmdaaGtkbmVIY2F6QUxBOFBpQ0E1cGJTWndLdlJMRGVYR1hwbCtT?=
=?utf-8?B?bEVVczVaTFRMVkhkZG4zT3pUN0orQUZnRWc0bXpZRW1jTEtqSTlZWGxhL08r?=
=?utf-8?B?RUpGTjU1SktaSHVqNURnUlJlejNSSmhpR3VlZWZrcEVtaFYwK2RqbGxsS0Ny?=
=?utf-8?B?MExubGorVitiZjhBd2FqS3FEMnhXNU01QUZsUjZLV1cvU2FXSEhORnRzUnc3?=
=?utf-8?B?T05IaFd5MnZrRElsSTBCWjAweUJlcUJ3MHN1ZVU5QUFxQnlIUjlabllqTmYw?=
=?utf-8?B?cDAyNkEvYmlKSkwyemR0YlYvUHMwV1AzZWptcEhWa1d6bFQ5RHF4UVI3SlhW?=
=?utf-8?B?cDFUdFQwRzg0QjIwNmFPTCtCMktmNHU0YzRwaHRDc0Iza000YTdYRTZFMWZh?=
=?utf-8?B?R0UzYnAzSWV3OWhyZ01GTmQ0OC94Q0c1RWxmNC9QUUlUZW4ydzVKWkh3enBj?=
=?utf-8?B?VzZhd3lkYnhYWHYveTdjbGQrWTU3S1ZWT2I5aVNMcXZwYjdYSThtYjhBVm1a?=
=?utf-8?B?eVBNeWlwOGJCRXphL0JsTm9jRXZ0LzByMmtRMklpNGFQcTB5cGI2L3h0T1lJ?=
=?utf-8?B?dDZhUUZwc011Q0F3bjMzcFd2dXN1bm1yUDQ2dEl2c1hnVzFvOTI3dE9JbkpW?=
=?utf-8?B?cFRvd0N2NG1rYytVeHVvRythUHJWcmVvcW1aWFNCODlEZVlrd1lDQXk1YUxO?=
=?utf-8?B?RWJaZzFnYjdnUmNQOS9OcWZ0MFliU1ZwU2thMFZwRnVsRDQ5SDlaZjJPZmV3?=
=?utf-8?B?QnhDTkU5V1VLTFNHVWh4MEdJZlJhZHlzbzVlbFlFOTA0cjAxcUZqRXBsdU5L?=
=?utf-8?B?V2ZMMGlPK0NHRCtCKytGd0c5VHFad282Y1JiWW80NW1ETmQyRDFxcFMvRExv?=
=?utf-8?B?NUxNZGU2Y1ZKd2VPOGJUNFpOaUZlTnFoT2ZLazZEb2xuQ2RQZk44QUhycElV?=
=?utf-8?B?Rm1FaHIxcXo0SUpiZFRybFp0dHlHb3pXVVZObURBUEE4VXMyTk1ISW1wd1oy?=
=?utf-8?B?eTVyZU11YWNlcFRlbHhUaCt3TUt1dUd2TUY2bzVHMTFJRUowZnM1Z080WHRF?=
=?utf-8?B?Qi9BTGZRcFlXS0RoUU15V0g5c3BJcHMvWVVKekhTRlpoNFExUFdkaGd2cDM1?=
=?utf-8?B?YkNYTjF2emVFeC8xTjg2cGlPdXgzdWxnVHhYWnhSSkJwbkQ4dkZUNzZvaUJK?=
=?utf-8?B?cUdYU1Z5QXVUYmp6Tk9tU25WZlhOblFXNFBsQlV5REFPTDZLUm40allobmZt?=
=?utf-8?B?VTl4Uk4vMnhvWkU4L05zVTVBMUhXcWRFN2ZlS3U2V2xsRm45TnNDUEJ2VWFN?=
=?utf-8?B?dis3Vlh0Rm0rR0krMGFHcWZraHIreE01Qkx3blJaNTJ4dUg1TDE3a1ZUeVgy?=
=?utf-8?B?SWVTNzdkSEhsNDBkNDhsQnFvL1F5M2syeGNEUnd5T29udUFOUDNCMmhyMzY0?=
=?utf-8?B?UzlEemtGUG10Y3FnSW9ER1FpZm1kV3JjWU5sL2YwWXdKaFh2WE1tQVRwQXZt?=
=?utf-8?B?MUZYS3lMK0lSVmNYbG41L1UzT0crZjVsWFo0ZVN4cWFRMm14NXlGVUpTSFFl?=
=?utf-8?Q?pBxY=3D?=
x-ms-exchange-antispam-messagedata-chunkcount: 1
x-ms-exchange-antispam-messagedata-0:
=?utf-8?B?N2ZvN0NhTVVhQVlqUzkvWW5ra0JnMzVKTUlXMkhOVmxTSzJmYjRxaXkzYm4y?=
=?utf-8?B?dDMzQ0kxTHR3UVpRSncvVENsWDhnSlJTa3VZNEREU0VCRmdyZkRRWTBnNlFX?=
=?utf-8?B?ck43UDlNYUVNdG5zUC8wcEpDV0NJU2wyam9zeVZoQ1dIWjhLQVMwZUhQTlFZ?=
=?utf-8?B?cW1zZldvWHFzalVSNFdwQmN5VjdOV2NMS2JJL1BWZUoxRXJXSHRVcjRNTWp5?=
=?utf-8?B?Q0lxcy9VaTRBQVQ0TlkwaHRkalFHSWtmWnB5cGtuS0pJdmJ1WjJPbExWc1d0?=
=?utf-8?B?Mjc3NmdNZW04aDlaR0hmeTZ6Y1dVUGJWUUNSS1hBSWphSWw0Z0dHZGloMzZn?=
=?utf-8?B?b1lOUWQ4Y3ZNY3NTYWFRT1h2UmF5VXBBZldqSHozTWlNODh3N2Ivc2VlV3Ey?=
=?utf-8?B?TlFIM2kvSVdsVlVzRGVvQzZDYk54bDlkejIxbVprMm9EUWRDd0tMTkdtRis2?=
=?utf-8?B?WEhBdkUwOVlNR3dVd2xMeFdHOE8wVkRuTnIrN09iQWc1YURmM2hHZVlWNnZS?=
=?utf-8?B?Y2NKUkFRYmE0UWVkYzl4NG8zbzAyV0hBd2I2VHdnNDhSNlBrMkg2dlFqdDNl?=
=?utf-8?B?cHVSK2ZhbnRkajl2NE13a2dIWUpEdjk0SzFTSWxCOWJxWTIvZ0RjVmVtWHJn?=
=?utf-8?B?RWZ3QzBqWXpvc2JVK0xzK2E5TndUQmZCY0sreEhwc1RvaGpHUVU4dlhtMENZ?=
=?utf-8?B?cjRuaHk2UTUxbnVLcDB6UDZmcURwRUJlcHJMWGJWOW5Rc2JhVFNGeDFld09J?=
=?utf-8?B?dTU3Y3hrUURxS040aXpNY1Radk1BMnNmRDh3TC9Ka1VpSDZ0V3NXYlFSalF6?=
=?utf-8?B?b2dTN3RGWGdReHNQck85Rk9EQ3k3VEMwUVBuVDZCTUFnbjFUVmR4VjhqUHZO?=
=?utf-8?B?NUNEK1liN2NVQVZvRWM5MVdUSk5lbjdTZ053Q1lyVW9CQkl4NVRkckgwZ2M1?=
=?utf-8?B?V21GS1I5MGpodVdvVXhvSWZHZ2hZOGhaNklIR0hTazdoTnhXaVh3N1RXeTRP?=
=?utf-8?B?R3RtRGVWNk1VeGY4MytWNjlCek1tWnEzekZIZy9sc1Aydk56MXRDNTRvOUhs?=
=?utf-8?B?NDA3S2lzUGtrR2s1RUs2cG9Vc25KeEkyN2l2SEx5Q24xbGhtYUhMZ0JobitD?=
=?utf-8?B?NzZDL3lubVBSRGZMeHBJK3EyU0taQUU5Q3YrT3NwMTUySmRrQ0xBSXZ3ZHdv?=
=?utf-8?B?clRxUDIwTVVEdTNUaTBSZG9ISmFaM01DRk5zTEluaGR3VThTWklBSS9qNTZL?=
=?utf-8?B?eGpZb0NnL0RzRFVvR2ZLYkNSMXVXSzVIVjdCVi9PelJ4NVdwNmE0UTV0ZmtW?=
=?utf-8?B?OTVGcERTaVZNZ0hCYWlac2Y0Z1FjbU91SEJuaTBqNURvVDZjMWk3ckoxQWM3?=
=?utf-8?B?cytWVWJYQ3prT1ppM0Q0RFBIbW9lZ3RQSzRtUWRVTlRoRDJoK0NJbVRub1dh?=
=?utf-8?B?ZzBsODdVK29lYUUyNEwxeVIzK1RlTE84aFp2OFdNNjlaOERBN2ZXdzRDR25T?=
=?utf-8?B?U29IaUIrRjVucmt2eUpCemlWSWZ2MEdobDZzazJNTytmaitKRyszSDNMb2Fl?=
=?utf-8?B?LzlNOWo2RElsQVdhV2VYY2xHd0lkT2FwZ21jRkFtQlU5ZHBhUit0VGdocEh2?=
=?utf-8?B?NXdtbDhocjFKOTJTMUQ5bDRlN3lKUDkwYjl1R0V1Y0ZJUnJ3UnRiSEtPNWNU?=
=?utf-8?B?b2JSUGN4OHVPUUFGSkY4TW5wa3M0enRwM0VCUXU1ZmNkZmJwV1ZtR3RhNmFU?=
=?utf-8?B?cHZlVXhrVXVYcTA3WUJkTjNNVVVCQ3ROaDU0KzB3UURhQ2MyNmNtN3lZYWFy?=
=?utf-8?B?eFBmMlJRN2UwQUtsR090ZTB1eEthemlEYVpIMWZuZEN5MGM4b0NKREkzTWJ1?=
=?utf-8?B?c0hyRUNpZmFBRUV3TWhmR3k0SFlEOThZb0oyWlVPRnUrZjVXVU4xd04wdCtH?=
=?utf-8?B?NmpkQlpYbHhkRThkNlhabmgxSmVWM2wzc3BQMzNvbmxSYVRuT2FyZUljK2ow?=
=?utf-8?B?S055bk4wUm5ycGFPSUVDTDBtQnNBTWdLRFRsQ2NSUWhick9lOFhmWlRock5j?=
=?utf-8?B?dXVFMUJwQm1HRm8zTDlYZC9YNXZ4Mk5iZ01sWlgwc0tncjYvMEt2NEpUcUEr?=
=?utf-8?B?VlF6T3hCWFB6UlJGUExlbXVrUTRGR0xCY0xaaldMWXRTR3JLTDhyWmRIVE9Z?=
=?utf-8?B?NU5PVzhycU9vd3dWNHc1S1ZUM3UyWkd2RUlTaGhlZXZCZjFjZ1JScXJnZmYv?=
=?utf-8?B?YWFCWlZPSFh0dlpnbVVvK3oza0lCeTFGdjg1TmZOcTBnQXYveFFVNTRRWVE0?=
=?utf-8?B?S2tLYXBSUk5FeGo5QUduLzR3UTMwV24rVExWZUwrTGVYZmhrNTYrUT09?=
X-OriginatorOrg: kent.ac.uk
X-MS-Exchange-CrossTenant-AuthAs: Internal
X-MS-Exchange-CrossTenant-AuthSource: AM6PR01MB5384.eurprd01.prod.exchangelabs.com
X-MS-Exchange-CrossTenant-Network-Message-Id: fc0db64e-eba3-46b9-99ff-08de6f9c0236
X-MS-Exchange-CrossTenant-originalarrivaltime: 19 Feb 2026 09:48:17.4841
(UTC)
X-MS-Exchange-CrossTenant-fromentityheader: Hosted
X-MS-Exchange-CrossTenant-id: 51a9fa56-3f32-449a-a721-3e3f49aa5e9a
X-MS-Exchange-CrossTenant-mailboxtype: HOSTED
X-MS-Exchange-CrossTenant-userprincipalname: l/JKFLEAEpc1HYYfi9OZRoteO24KCIhke+IKWvdgfTTUJnXjUtDHaWRYvFKNVDBh
X-MS-Exchange-Transport-CrossTenantHeadersStamped: GV1PR01MB10739
Precedence: Bulk
List-Subscribe: <mailto:main+subscribe_at_ammrl.groups.io>
List-Help: <mailto:main+help_at_ammrl.groups.io>
Sender: main_at_ammrl.groups.io
List-Id: <main.ammrl.groups.io>
Mailing-List: list main_at_ammrl.groups.io; contact main+owner_at_ammrl.groups.io
Delivered-To: mailing list main_at_ammrl.groups.io
Resent-Date: Thu, 19 Feb 2026 03:51:21 -0800
Resent-From: G.S.Thompson_at_kent.ac.uk
Reply-To: main_at_ammrl.groups.io
List-Unsubscribe-Post: List-Unsubscribe=One-Click
List-Unsubscribe: <https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/plugh>
X-Gm-Message-State: Bkpaw20aOdPpNVPCeaGXoZR6x7559972AA=
Content-Language: en-US
Content-Type: multipart/mixed;
boundary="_006_447997EC3020490DA99FA8622A1796CAkentacuk_"
X-Proofpoint-GUID: KyvTshxipgfkl8LfF9P4hiANmo2P75ru
X-Authority-Analysis: v=2.4 cv=IocTsb/g c=1 sm=1 tr=0 ts=6996f93c b=1 cx=c_pps
a=7zJfR2BOg5ERPHYvypTuCg==:117 a=7zJfR2BOg5ERPHYvypTuCg==:17
a=z/mQ4Ysz8XfWz/Q5cLBRGdckG28=:19 a=HzLeVaNsDn8A:10 a=FxERRFPGSi8A:10
a=SKaEv8EmnvEA:10 a=ZKS8Kw3IVw4A:10 a=Kan8h273Bse8B8Jy5C1n:22
a=Mpw57Om8IfrbqaoTuvik:22 a=GgsMoib0sEa3-_RKJdDe:22 a=XYAwZIGsAAAA:8
a=JOA0iUbgAAAA:8 a=5cl2wky9AAAA:8 a=xjGz-wBMAAAA:8 a=Gh3JI2XeAAAA:8
a=H_CbGO1CAAAA:8 a=RApa2X0Ju_9npLaaNNgA:9 a=QEXdDO2ut3YA:10 a=SSmOFEACAAAA:8
a=t94Cn9Xn7_wMG88EmC8A:9 a=44nEihtXJ3dxEuTI:21 a=_W_S_7VecoQA:10
a=lqcHg5cX4UMA:10 a=HXjIzolwW10A:10 a=T6a71-JsGAwA:10
a=VI9WJXAUdhXo16XYtocA:9 a=UO-yzy6s3cQA:10 a=I2lQ5O3grGkA:10
a=E8ToXWR_bxluHZ7gmE-Z:22 a=nG9pSErA777cby-CJFl-:22 a=l6ZfdfcawmuAxJFV5O2W:22
a=dmbvPvuLAiAwK5y7oBu2:22 a=SRMwOmkBraWqXViabrVv:22 cc=dsc
X-CLX-Response: 1TFkXGBkSEQpMehcZHx4RCllEF219U0JOTWtvHV9nEQpYWBdubl1TGmdCbmR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 g
X-Proofpoint-ORIG-GUID: KyvTshxipgfkl8LfF9P4hiANmo2P75ru
X-CLX-Shades: MLX
X-Proofpoint-Spam-Details-Enc: AW1haW4tMjYwMjE5MDEwOCBTYWx0ZWRfX1hEZMQ46hQUG
SfLFroO7u6Y2uUiwvdwC+SxFTUYZBBGQzFWtJKeIJqNSJfsy3tGCYJPw6FrRRcJVgB0MPOartk0
xDU2MOIckvS7TqWfe13EyvwZmVgqsllT+4I/HYeLB9b5wOw2qaAvDVtQXXpSDxQ7fbS5+knuczW
sdpCVmQ08Mh9c0IcA4VwC9QGVHqbcuBvoUgBSTNDdXon8pl+gxQdRLOyEHsNpyk55OT4S4bAJRa
zbbc2Fi21GiZvtKIpo/JVL7rQXXS6gXiTZEQRGMurySBS427O62mO/Wh7SVC9x4ZavOBE46NZBE
RjEdacdK5Jph5e3ouOYyQwplBRzzkxrQOk3qBq/BeYi+HRxZLLdcn4fpqvoLsSd/2xxA6iF7cN7
D7UVg8uIZa1f05KS4Z7ZrDo/fGkNvtowKVSVMSbff9xwYB6ZEkjPTu6nRmz4NbXHUl/07/tOzkI
CrfAK14D9Ok5EYVdiW3WozBmYPVtZY4fqIibP2vExM6rncil+yENje8vEW9Xg02vZFD5oQgZkLm
XMZN/xAf1kxhyRsZoUBPmNOjFHltE=
MIME-Version: 1.0
X-Proofpoint-Virus-Version:
X-Proofpoint-Spam-Details: rule=inbound_notspam policy=inbound score=0
suspectscore=0 lowpriorityscore=0 spamscore=0 phishscore=0 malwarescore=0
priorityscore=354 impostorscore=0 adultscore=0 bulkscore=0 clxscore=238
classifier=typeunknown authscore=0 authtc= authcc=discussion route=inbound
adjust=0 reason=mlx scancount=1 engine=8.22.0-2601150000
definitions=main-2602190108 domainage_hfrom=5582 domainage_replyto=5582
--_006_447997EC3020490DA99FA8622A1796CAkentacuk_
Content-Type: multipart/related;
boundary="_005_447997EC3020490DA99FA8622A1796CAkentacuk_";
type="multipart/alternative"
--_005_447997EC3020490DA99FA8622A1796CAkentacuk_
Content-Type: multipart/alternative;
boundary="_000_447997EC3020490DA99FA8622A1796CAkentacuk_"
--_000_447997EC3020490DA99FA8622A1796CAkentacuk_
Content-Type: text/plain; charset="utf-8"
Content-Transfer-Encoding: quoted-printable
Looks like Vectron was bough by microchip.com
<
http://microchip.com >
<
https://www.microchip.com/en-us/about/corporate-overview/acquisitions/vectron >
Microchip Technology
https://www.microchip.com =E2=80=BA ... =E2=80=BA Acquired Companies
Find the Oscillators, Crystals and SAW filters offered by Vectron International.
Microchip acquired the Vectron product line with the acquisition of Microsemi =E2=80=A6
The OX-205 is in their catalogue, I would contact them for distributors
https://ww1.microchip.com/downloads/aemDocuments/documents/VOP/ProductDocuments/Brochures/VI-product_brochure.pdf
Mouser also seem to have them listed but you need to contact them
https://eu.mouser.com/ProductDetail/Microchip-Technology/OX-2050-BEE-2072-80M0?qs=3DvLWxofP3U2y4rUtwH8JoMw*3D*3D&srsltid=3DAfmBOoo6soe6lLIqKiongEEX30sY6YQYL1BzLx_D3mtzvF_VrJyRl1Ow
Or
https://www.directindustry.com/prod/vectron-international/product-34784-1541801.html
Say they will quote you a price
regards
Gary
Dr Gary S Thompson NMR Facility Manager
CCPN CoI & Working Group Member
Wellcome Trust Biomolecular NMR Facility
School of Natural Sciences
University of Kent, Canterbury, Kent, England, CT2 7NZ
=E2=98=8E:01227 82 7117
=E2=9C=89=EF=B8=8F: g.s.thompson_at_kent.ac.uk
orchid: orcid.org/0000-0001-9399-7636
On 18 Feb 2026, at 18:11, Kreutz, Christoph Roman via groups.io wrote:
Vectron
OX-204-0005
-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3245): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3245__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC4=
3U1xsDZXsMuZRL8X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUpWTMD3w$=20
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117874352=
/7559972__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC43U1xsDZXsMuZRL8=
X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUpGxDRwc$=20
Group Owner: main+owner_at_ammrl.groups.io
-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-
--_000_447997EC3020490DA99FA8622A1796CAkentacuk_
Content-Type: text/html; charset="utf-8"
Content-ID: <6795BF41C98ECB4FB948EF481147DA75_at_eurprd01.prod.exchangelabs.com>
Content-Transfer-Encoding: quoted-printable
<html aria-label=3D"message body"><head>
<meta http-equiv=3D"Content-Type" content=3D"text/html; charset=3Dutf-8"/>
</head>
<body style=3D"overflow-wrap: break-word; -webkit-nbsp-mode: space; lin=
e-break: after-white-space; ; ;">
Looks like Vectron was bough by <a href=3D"
https://urldefense.com/v3/__http=
://microchip.com__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC43U1xsDZ=
XsMuZRL8X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUoGpjwf4$">microchip.com</a>
<div><br/>
</div>
<div>
<div class=3D"kb0PBd A9Y9g jGGQ5e" data-snf=3D"x5WNvb" data-snhf=3D"0" styl=
e=3D"flex: 1 1 100%; min-width: 0px; caret-color: rgb(31, 31, 31); ; ;=
">
<div class=3D"yuRUbf">
<div class=3D"b8lM7" style=3D"display: flex; flex-direction: column-rever=
se; ; ;"><span class=3D"V9tjod" jsaction=3D"trigger.mLt3mc" style=3D"text-=
wrap: wrap; transform: scaleY(-1); ; ;"><a jsname=3D"UWckNb" class=3D"zR=
eHs" href=3D"
https://urldefense.com/v3/__https://www.microchip.com/en-us/ab=
out/corporate-overview/acquisitions/vectron__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBj=
xeqoB92fn3tvQVLlwzC43U1xsDZXsMuZRL8X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyU=
kOdF30z$" data-ved=3D"2ahUKEwiG_7WUouWSAxXWSEEAHWjNKbAQFnoECBoQAQ" data-sb=
=3D"/url?sa=3Dt&source=3Dweb&rct=3Dj&opi=3D89978449&url=3Dh=
ttps://www.microchip.com/en-us/about/corporate-overview/acquisitions/vectro=
n&ved=3D2ahUKEwiG_7WUouWSAxXWSEEAHWjNKbAQFnoECBoQAQ&usg=3DAOvVaw1C2=
M9cF-FwAqmg7axAlKjB" style=3D" text-decoration: none; outline: 0px; ; ;">
<div class=3D"notranslate ESMNde HGLrXd ojE3Fb" style=3D"display: inline-bl=
ock; padding: 0px; max-width: calc(100% - 20px); text-wrap: wrap; t=
ransform: scaleY(-1); ; ;">
<div class=3D"q0vns" style=3D"display: flex; padding: 0px; overflow: hi=
dden; align-items: center; ; ;">
<span class=3D"DDKf1c" style=3D"margin-right: 12px; ; ;">
<div class=3D"eqA2re UnOTSe Vwoesf" aria-hidden=3D"true" style=3D"vertical-=
align: middle; display: inline-block; ; ;">
<img class=3D"XNo5Ab" src=3D"cid:291550EF-73EA-4F7C-AF80-FC49B9019DF9" alt=
=3D"gAAAABJRU5ErkJggg=3D=3D.png" data-csiid=3D"ONuWaYaDKdaRhbIP6JqngQs_4" d=
ata-atf=3D"1" style=3D"border: 1px solid var(--mXZkqc); border-radius: 50%;=
display: block; ; " width=3D"28" height=3D"28"/></div>
</span>
<div class=3D"CA5RN" style=3D"overflow: hidden; ; ;">
<div><span class=3D"VuuXrf" style=3D" display: block; line-height: 20px; =
white-space: nowrap; overflow-x: hidden; text-overflow: ellipsis; w=
idth: auto; ; ;">Microchip Technology</span></div>
<div class=3D"byrV5b" style=3D"align-items: center; display: flex; flex=
-direction: row; ; ;">
<cite class=3D"qLRx3b tjvcx GvPZzd cHaqb" role=3D"text" style=3D" overflow:=
hidden; text-overflow: ellipsis; white-space: nowrap; line-height: 1=
8px; max-width: max-content; width: 290.496307px; ; ;">https://www.mic=
rochip.com<span class=3D"ylgVCe ob9lvb" role=3D"text">=C2=A0=E2=80=BA
... =E2=80=BA Acquired Companies</span></cite></div>
</div>
</div>
</div>
</a></span>
<div class=3D"B6fmyf byrV5b Mg1HEd" style=3D"align-items: center; display=
: flex; flex-direction: row; max-height: 0px; position: relative; t=
op: 19px; visibility: hidden; height: auto; ; ;">
<div class=3D"HGLrXd ojE3Fb" style=3D"display: inline-block; padding: 0px=
; max-width: calc(100% - 20px); ; ;">
<div class=3D"q0vns" style=3D"display: flex; padding: 0px; overflow: hi=
dden; align-items: center; ; ;">
<span class=3D"DDKf1c" style=3D"margin-right: 12px; ; ;">
<div class=3D"eqA2re UnOTSe" style=3D"height: 26px; width: 26px; ; ;"></=
div>
</span>
<div class=3D"CA5RN" style=3D"overflow: hidden; ; ;">
<div><span class=3D"VuuXrf" style=3D" display: block; line-height: 20px; =
white-space: nowrap; overflow-x: hidden; text-overflow: ellipsis; w=
idth: auto; ; ;"></span></div>
<div class=3D"byrV5b" style=3D"align-items: center; display: flex; flex=
-direction: row; ; ;">
<cite class=3D"qLRx3b tjvcx GvPZzd cHaqb" role=3D"text" style=3D" overflow:=
hidden; text-overflow: ellipsis; white-space: nowrap; line-height: 1=
8px; max-width: max-content; width: 290.496307px; ; ;"><span class=3D"=
ylgVCe ob9lvb" role=3D"text"></span></cite></div>
</div>
</div>
</div>
<div class=3D"csDOgf BCF2pd ezY6nb L48a4c" style=3D"display: inline; visi=
bility: visible; margin-left: 8px; position: relative; height: 18px; =
margin-top: 16px; ; ;">
<div jsdata=3D"l7Bhpb;_;ONuWaYaDKdaRhbIP6JqngQs18" jscontroller=3D"i8S0p" i=
d=3D"atritem-_ONuWaYaDKdaRhbIP6JqngQs_54" jsslot=3D"" jsaction=3D"rcuQ6b:np=
T2md;h5M12e" data-ved=3D"2ahUKEwiG_7WUouWSAxXWSEEAHWjNKbAQ2esEegQIGhAJ">
<div class=3D"MJ8UF iTPLzd rNSxBe eY4mx lUn2nc" aria-describedby=3D"_ONuWaY=
aDKdaRhbIP6JqngQs_44" aria-label=3D"About this result" role=3D"button" tabi=
ndex=3D"0" style=3D"padding-right: 0px; padding-left: 0px; padding-bott=
om: 20px; cursor: pointer; top: 0px; line-height: 16px; left: 0px; =
width: 28px; z-index: 1; position: absolute; outline: 0px; ; ;">
<span class=3D"D6lY4c" style=3D"height: 22px; width: 22px; position: ab=
solute; border-radius: 11px; left: 13.98897px; transform: translateX(=
-50%); text-align: center; top: -1px; ; ;"><span class=3D"xTFaxe z1asC=
e" style=3D"display: inline-block; fill: currentcolor; height: 18px; =
line-height: 18px; position: relative; width: 18px; top: 3px; ; ;"><=
svg focusable=3D"false" aria-hidden=3D"true" xmlns=3D"
http://www.w3.org/200=
0/svg" viewBox=3D"0 0 24 24"><path d=3D"M12 8c1.1 0 2-.9 2-2s-.9-2-2-2-2 .9=
-2 2 .9 2 2 2zm0 2c-1.1 0-2 .9-2 2s.9 2 2 2 2-.9 2-2-.9-2-2-2zm0 6c-1.1 0-2=
.9-2 2s.9 2 2 2 2-.9 2-2-.9-2-2-2z"></path></svg></span></span></div>
</div>
</div>
</div>
</div>
</div>
</div>
<div class=3D"kb0PBd A9Y9g" data-snf=3D"nke7rc" data-sncf=3D"1" style=3D"fl=
ex: 1 1 100%; min-width: 0px; caret-color: rgb(31, 31, 31); ; ;">
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
Find the Oscillators, Crystals and SAW filters offered by=C2=A0<span style=
=3D"">Vectron</span>=C2=A0International.=C2=A0<span style=3D"">Microchip</s=
pan>=C2=A0acquired the=C2=A0<span style=3D"">Vectron</span>=C2=A0product
line with the acquisition of Microsemi=C2=A0=E2=80=A6</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
The OX-205 is in their catalogue, I would contact them for distributors</di=
v>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<div style=3D"display: block; ; ;"><img width=3D"348" alt=3D"PastedGraphic=
-1.tiff" src=3D"cid:7E6676E7-D8C0-4192-9FB1-F88C85809069" height=3D"648"/><=
a href=3D"
https://urldefense.com/v3/__https://ww1.microchip.com/downloads/a=
emDocuments/documents/VOP/ProductDocuments/Brochures/VI-product_brochure.pd=
f__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC43U1xsDZXsMuZRL8X0MvLvW=
AE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUoXKdbEz$">
https://ww1.microchip.com/downlo=
ads/aemDocuments/documents/VOP/ProductDocuments/Brochures/VI-product_brochu=
re.pdf</a></div>
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
Mouser also seem to have them listed but you need to contact them</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<a href=3D"
https://urldefense.com/v3/__https://eu.mouser.com/ProductDetail/=
Microchip-Technology/OX-2050-BEE-2072-80M0?qs=3DvLWxofP3U2y4rUtwH8JoMw*3D*3=
D&srsltid=3DAfmBOoo6soe6lLIqKiongEEX30sY6YQYL1BzLx_D3mtzvF_VrJyRl1Ow__;=
JSU!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC43U1xsDZXsMuZRL8X0MvLvWA=
E_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUkgv8hjP$">
https://eu.mouser.com/ProductDeta=
il/Microchip-Technology/OX-2050-BEE-2072-80M0?qs=3DvLWxofP3U2y4rUtwH8JoMw%3=
D%3D&srsltid=3DAfmBOoo6soe6lLIqKiongEEX30sY6YQYL1BzLx_D3mtzvF_VrJyRl1Ow=
</a></div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
Or=C2=A0</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<a href=3D"
https://urldefense.com/v3/__https://www.directindustry.com/prod/=
vectron-international/product-34784-1541801.html__;!!PvDODwlR4mBZyAb0!WZvZG=
Z3MBjxeqoB92fn3tvQVLlwzC43U1xsDZXsMuZRL8X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4=
GnpyUoQVSqLA$">
https://www.directindustry.com/prod/vectron-international/pr=
oduct-34784-1541801.html</a></div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
Say they will quote you a price</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
regards</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
Gary</div>
<div class=3D"VwiC3b yXK7lf p4wth r025kc Hdw6tb" style=3D" line-height: var=
(--XWv1if); word-break: break-word; display: -webkit-box; -webkit-box=
-orient: vertical; overflow: hidden; -webkit-line-clamp: 2; ; ;">
<br/>
</div>
</div>
<div>
<div dir=3D"auto" style=3D"text-align: start; text-indent: 0px; overflo=
w-wrap: break-word; -webkit-nbsp-mode: space; line-break: after-white-s=
pace; ; ;">
<div style=3D"caret-color: rgb(0, 0, 0); letter-spacing: normal; text-t=
ransform: none; white-space: normal; word-spacing: 0px; text-decorati=
on: none; -webkit-text-stroke-width: 0px; ; ;">
Dr Gary S Thompson NMR Facility=C2=A0Manager<br/>
CCPN CoI & Working Group Member<br/>
Wellcome Trust Biomolecular NMR Facility=C2=A0=C2=A0<br/>
School of Natural=C2=A0Sciences=C2=A0<br/>
University of Kent,=C2=A0=C2=A0Canterbury,=C2=A0=C2=A0Kent,=C2=A0=C2=A0Engl=
and,=C2=A0=C2=A0CT2 7NZ=C2=A0=C2=A0<br/>
=C2=A0<br/>
=E2=98=8E:01227 82 7117=C2=A0=C2=A0<br/>
=E2=9C=89=EF=B8=8F: g.s.thompson_at_kent.ac.uk<br/>
orchid: orcid.org/0000-0001-9399-7636</div>
<div></div>
</div>
<br class=3D"Apple-interchange-newline"/>
<br class=3D"Apple-interchange-newline"/>
</div>
<div><br/>
<blockquote type=3D"cite">
<div>On 18 Feb 2026, at 18:11, Kreutz, Christoph Roman via groups.io <Ch=
ristoph.Kreutz=3Duibk.ac.at_at_groups.io> wrote:</div>
<br class=3D"Apple-interchange-newline"/>
<div>
<div style=3D"caret-color: rgb(0, 0, 0); font-variant-caps: normal; let=
ter-spacing: normal; text-align: start; text-indent: 0px; text-transf=
orm: none; white-space: normal; word-spacing: 0px; -webkit-text-strok=
e-width: 0px; text-decoration: none; ; ;">
Vectron=C2=A0</div>
<div style=3D"caret-color: rgb(0, 0, 0); font-variant-caps: normal; let=
ter-spacing: normal; text-align: start; text-indent: 0px; text-transf=
orm: none; white-space: normal; word-spacing: 0px; -webkit-text-strok=
e-width: 0px; text-decoration: none; ; ;">
OX-204-0005</div>
</div>
</blockquote>
</div>
<br/>
</div>
</body></html>
<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>
<hr>
Groups.io Links:<p>
=20
You receive all messages sent to this group.
=20
=20
<p>
<a target=3D"_blank" href=3D"
https://urldefense.com/v3/__https://ammrl.grou=
ps.io/g/main/message/3245__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwz=
C43U1xsDZXsMuZRL8X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUpWTMD3w$">View/Rep=
ly Online (#3245)</a> |
<a target=3D"_blank" href=3D"mailto:main_at_ammrl.groups.io?subject=3DRe:%20=
Re%3A%20%5BAMMRL%5D%20OCXO%20REF1200%20board">Reply to Group</a>
=20
| <a target=3D"_blank" href=3D"mailto:G.S.Thompson_at_kent.ac.uk?subject=
=3DPrivate:%20Re:%20Re%3A%20%5BAMMRL%5D%20OCXO%20REF1200%20board">Reply to =
Sender</a>
=20
|
<a target=3D"_blank" href=3D"
https://urldefense.com/v3/__https://groups.i=
o/mt/117874352/7559972__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC43=
U1xsDZXsMuZRL8X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUpGxDRwc$">Mute This T=
opic</a>
| <a href=3D"
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/pos=
t__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC43U1xsDZXsMuZRL8X0MvLvW=
AE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUg4ZM3Gu$">New Topic</a>
<br>
<a href=3D"
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/edits=
ub/7559972__;!!PvDODwlR4mBZyAb0!WZvZGZ3MBjxeqoB92fn3tvQVLlwzC43U1xsDZXsMuZR=
L8X0MvLvWAE_IIqFmmRhQJJ02ok2j6hc6w8H4GnpyUsUwQyc3$">Your Subscription</a> |
<a href=3D"mailto:main+owner_at_ammrl.groups.io">Contact Group Owner</a> |
<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>
--_000_447997EC3020490DA99FA8622A1796CAkentacuk_--
Received on Wed Feb 18 2026 - 14:33:22 MST