[AMMRL] OCXO REF1200 board

From: Kreutz, Christoph Roman via groups.io <Christoph.Kreutz=uibk.ac.at_at_groups.io>
Date: Wed, 18 Feb 2026 07:29:25 +0000

Dear all,

We have a failed AV4 REF1200 ECL01 board in a 2017 Avance Neo console. Our
diagnostics indicate that the oven-controlled crystal oscillator (OCXO) is
defective. When we supply a 10 MHz sine-wave reference from an external signal
generator via the J1 10 MHz remote input, the board operates normally again.

Does anyone know of a commercial supplier for the OCXO used on the REF1200
board, or an equivalent replacement?

Thank you in advance for your help.

Best regards,
Christoph

---