Re: [AMMRL] SOP for Methlene Chloride NMR Sample prep.

From: Teipel, Jan (CVUA-KA) via groups.io <"Teipel,>
Date: Thu, 12 Feb 2026 18:37:03 +0000

Are these the „two-stepped” caps with two “ring
ribs” on the inside? They close the tube nice and tight and do not
pop off easily, but as far as I know they are still made of PP.

These really keep the solvent trapped inside:
https://www.sigmaaldrich.com/US/en/product/aldrich/z220612
https://sp-wilmadlabglass.com/product/5-mm-ptfe-nmr-cap-precision-tubes-only-25-pk/

Too expensive to be discarded after single use, but easily cleaned. And after
capping really ugly samples one can still bid them farewell.

It is good to have a few options on hand…

Kind regards / Freundliche Grüße

Jan TEIPEL (he/him)
CVUA Karlsruhe
Laborleiter NMR – Laboratory Head NMR
Obmann / Convenor DIN & CEN work groups
“NMR for Food Authenticity”

Weißenburger Straße 3
76187 Karlsruhe

Telefon: (+49) 721 926-3641
E-Mail: Jan.Teipel_at_cvuaka.bwl.de
Internet: http://www.ua-bw.de

(CVUA-KA, the State Institute for Chemical Analysis and Veterinary Diagnostics
Karlsruhe is an ancillary to the ministry for rural affairs, nutrition and
consumer protection of Baden-Wuerttemberg)



> From: main_at_ammrl.groups.io On Behalf Of Olson, Dean via groups.io
> Sent: Thursday, February 12, 2026 5:19 PM
> To: main_at_ammrl.groups.io
> Subject: EXTERN: Re: [AMMRL] SOP for Methlene Chloride NMR Sample prep.


We use a special NMR tube cap for DCM samples. Among other things, this
seems to make our safety people happy.

https://shop.nmrtubes.com/caps/norloc-security-caps

See attached solvent loss data from the maker.

***********************************
Dean L. Olson, Ph.D.
Director, NMR & EPR Labs
146 Roger Adams Lab
School of Chemical Sciences
U. of Illinois at Urbana-Champaign
600 S. Mathews Avenue
Urbana, IL 61801 USA
Lab: 217-244-0564; Cell: 217-722-9432
dolson_at_illinois.edu
http://scs.illinois.edu/nmr
http://scs.illinois.edu/epr/
***********************************

> From: main_at_ammrl.groups.io On Behalf Of Teipel, Jan (CVUA-KA) via groups.io
> Sent: Thursday, February 12, 2026 3:56 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] SOP for Methlene Chloride NMR Sample prep.

Dear Perry,

I found an SOP published at the Free University of Berlin (Germany):
https://www.bcp.fu-berlin.de/chemie/chemie/forschung/InorgChem/agspandl/Ressourcen/Betriebsanweisungen-Chemikalien-Praktikum/Betriebsanweisung-Dichlormethan.pdf

====
Hazardous Substance Identification
Dichloromethane, CH₂Cl₂
as a solvent for analytical and preparative purposes
GHS Health hazard symbol (torso with “star”)
GHS Warning symbol (exclamation mark)
Signal word: Danger

  * H315: Causes skin irritation.
  * H319: Causes serious eye irritation.
  * H335: May cause respiratory irritation.
  * H336: May cause drowsiness or dizziness.
  * H351: Suspected of causing cancer.
  * H373: May cause damage to organs through prolonged or repeated exposure.
Affected organs: Liver, blood, central nervous system
________________________________
Hazards to Humans and the Environment
________________________________
Protective Measures and Rules of Conduct

  * Do not inhale vapours.
  * Avoid contact with eyes, skin, mucous membranes, and clothing.
  * Do not spill.
  * Keep container tightly closed.
  * Keep away from sources of ignition.
  * Work only in a fume hood. Keep the front sash closed.

________________________________
Measures in Case of Danger

  * In case of spillage, cordon off the affected area, evacuate all persons
  in the room and notify the responsible laboratory manager immediately.
(My comment: For minor spillage, absorbing material is not recommended, just
let the spillage evaporate, milliliters of spillage inside the fume hood do
not require cordoning off and people evacuation)

________________________________
First Aid

  * After skin contact: Wash with plenty of water.
  * After eye contact: Rinse with plenty of water for at least 15 minutes
 with eyelids held open.
  * After inhalation: Move to fresh air immediately. Seek medical attention.
  * After ingestion: Seek medical attention immediately.
  * After contact with clothing: Remove contaminated or soaked clothing
(including undergarments) and personal protective equipment immediately.

________________________________
Proper Disposal

  * Do not dispose of in sink drain (waste water sewer)
  * Dispose of as a halogenated organic solvent in the fume hood (labelled container).
  * See operating instruction for waste disposal



The information seems complete and correct and it is in accordance with the
data offered by GESTIS (https://www.dguv.de/ifa/gestis/gestis-stoffdatenbank/index-2.jsp)
a database under oversight of the Germany Social Accident Insurance, we use this regularly
for our workplace safety solutions. GESTIS is available in English as well.

As NMR lab manager and work safety counselor of our institute I would add t=
he following rules of conduct:

Store in a ventilated closet

Use protective gloves made from fluorocarbon rubber (FKM). (Do not use nitrile
gloves) (my comment: Only for handling of larger volumes than NMR tube fillings)

Do not bring in contact with aluminium

Close filled NMR tubes with PTFE stoppers/ caps
(My comment: PTFE caps effectively stop evaporation over weeks and can be washed and recycled)

I used Chat GPT to translate the text of the SOP into English, to me, the translation
seems correct (my English proficiency is C1/C2, quite good). Naturally you will still
have to edit and format the text according to your institution’s SOP document style.

I hope this helps.

Kind regards

Jan Teipel


Jan TEIPEL (♂)
CVUA Karlsruhe
Laborleiter NMR – Laboratory Head NMR
Obmann / Convenor DIN & CEN work groups
“NMR for Food Authenticity”

Weißenburger Straße 3
76187 Karlsruhe

> From: main_at_ammrl.groups.io On Behalf Of Perry Pellechia via groups.io
> Sent: Monday, February 9, 2026 9:20 PM
> To: main_at_ammrl.groups.io<mailto:main_at_ammrl.groups.io>
> Subject: EXTERN: [AMMRL] SOP for Methlene Chloride NMR Sample prep.


Hello All,

Our Health and Safety lab is requiring us to have a specific SOP for using
deuterated methylene chloride as an NMR solvent. We do not use the solvent
very often and probably can live without it but we have a few dozen grams
and it's fairly expensive. I am hesitant (too cheap) to discard it as hazardous
waste.

I am trying to find a template SOP for using it prepare NMR samples so I can
edit for my need. Every example I can find is for usage at higher volumes.
Does anyone have an SOP for methylene chloride that can help me satisfy our
H&S people?

Thanks,
--
Perry.
--

Perry J. Pellechia
Director of NMR Services
Research Professor
University of South Carolina
Department of Chemistry and Biochemistry



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3224): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3224__;!!PvDODwlR4mBZyAb0!XlhRXlBXWz1JITnmUDlm7ht-NaAm2i=
awOTKLJFHSRcvb0FLKvPVj1RXbGTFOgCs1p5F6zsnIbU9wS7fI2FzyQPFLx82m$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117727044=
/7559972__;!!PvDODwlR4mBZyAb0!XlhRXlBXWz1JITnmUDlm7ht-NaAm2iawOTKLJFHSRcvb0=
FLKvPVj1RXbGTFOgCs1p5F6zsnIbU9wS7fI2FzyQLVPCA-k$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Feb 12 2026 - 12:23:07 MST

This archive was generated by hypermail 2.4.0 : Sun Mar 01 2026 - 13:47:31 MST