Holger,

You should check with Bruker about your specific hardware. There are various
hardware combinations that have 6Li in the 2H filter region, at the preamp
or in the probe (e.g., Prodigy probes; 17O may fall into the same on a Prodigy
probe). I believe the "standard" configuration on some consoles has this
issue, and you have to request special configurations during purchase so
6Li can be done. Bruker should be able to assist you in determining the
situation for your hardware.

Charlie

________________________________
> From: main_at_ammrl.groups.io on behalf of Frank Holger Foersterling via groups.io
> Sent: Wednesday, February 11, 2026 1:29 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Adding new nucleus to prosol/peakpowers

Hello everybody,

I am trying to run a new nucleus (6Li) for the first time on a Bruker Avance
III HD, running Topspin 3.8. I notice that there is no entry for that nucleus
in either the PROSOL (edprosol) or PeakPower table (edprobe). I tried adding
an entry in the PeakPower list, but the nucleus did not show up as an
option in the pulldown menu, and there seems to be no way to add another
nucleus in thje PROSOL table.

I was able to copy a CORETAB entry (17O) to 6Li in TS/conf/instr/spec/cortab,
which is recognized in cortab, but it does not seem to add that nucleus as
an option to the PROSOL table.

It does show up in ednuc as a known nucleus.

I was able to get entries for 7Li into both PeakPower and Prosol, but some
references indicate that 6Li may give sharper lines sometimes, so I was
going to give that a try also.

Thanks for any advise

Holger Foersterling




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3215): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3215__;!!PvDODwlR4mBZyAb0!Tpt0KRbTLGBxKxkuLQ1Cl9OrenOk5I=
asYtFDKhKFtKamFKmh5J98wAV42yp8F193mxAchInRjZUu2MCWUXxWUiVLEA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117761768=
/7559972__;!!PvDODwlR4mBZyAb0!Tpt0KRbTLGBxKxkuLQ1Cl9OrenOk5IasYtFDKhKFtKamF=
Kmh5J98wAV42yp8F193mxAchInRjZUu2MCWUXzgKfzQdQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Feb 11 2026 - 11:48:34 MST

This archive was generated by hypermail 2.4.0 : Sun Mar 01 2026 - 13:47:31 MST