Hello everybody,
I am trying to run a new nucleus (6Li) for the first time on a Bruker Avance III HD,
running Topspin 3.8. I notice that there is no entry for that nucleus in either the
PROSOL (edprosol) or PeakPower table (edprobe). I tried adding an entry in the
PeakPower list, but the nucleus did not show up as an option in the pulldown menu,
and there seems to be no way to add another nucleus in thje PROSOL table.
I was able to copy a CORETAB entry (17O) to 6Li in TS/conf/instr/spec/cortab, which
is recognized in cortab, but it does not seem to add that nucleus as an option to
the PROSOL table.
It does show up in ednuc as a known nucleus.
I was able to get entries for 7Li into both PeakPower and Prosol, but some references
indicate that 6Li may give sharper lines sometimes, so I was going to give that a try also.
Thanks for any advise
Holger Foersterling
-=-=-=-=-=-=-=-=-=-=-=-