As others have mentioned, the best solution is to add an air dryer, prior
to the BCU or after the -40C air dryer.
BRUKER, does offer one that can take -40C air to -80C.
It was quoted at around 5k Euro a couple of years ago. It was described
like this
AH0093
VT GAS Membrane Air Dryer (HR + DVT)
Membrane air dryer for SmartCoolers (BCU-I, BCU-II).
Can be used together with High Resolution NMR liquids, DVT-style MAS & new
DVT-style
HR-MAS probes.
Features:
- max. flow 60 liter/min
- dew point -80 °C
# For WVT-style MAS probes use AH0083 instead
# May also be used together with BCU-05 and BCU-X
Hope it helps.
Cheers
On Fri, 23 Feb 2024, 19:59 Xia, Youlin via groups.io, wrote:
> Dear all,
>
> We are considering purchasing an air compressor working for both 400 MHz
> (room temp probe with BCU1; ordered) and 700 MHz (cryoprobe with BCU2;
> currently working with a small air compressor temporarily). From Bruker’s
> manual, BCU1 requires a dew point of -50C for the compressed gas and BCU2
> requires -80C. Now the situation we are facing is the air compressor with
> dew point of -70C is much expensive and with several months lead time. The
> one with the dew point of -40C is available and cheaper. We hope to get the
> 400 MHz installed as early as possible and can’t wait the one with -70C
> ready.
>
> We really appreciate you can provide us your experience and advice on
> choosing the right air compressor.
>
> 1, Does the air compressor with dew point of -40C work well for your
> machines with BCU2?
>
> 2, Is a dryer a must have if you use the air compressor with dew point of
> -40C? If yes, could you please provide the details (brand, model, cost,
> reliability, etc.)?
>
> 3, Do you think it must use the air compressor with dew point of -70C for
> NMR machines with BCU2 (probable we only found one type with -70C)?
>
> 4, Could you please recommend the air compressor with dew point lower than
> -70C if we must to have?
>
> 5, Any other advice for the compressed air supply with such required
> quality for our case (equipped with BCU1 and BCU2)?
>
>
>
> Really appreciate your information!
>
> Thanks,
>
> Fahu
>
> Fahu.he_at_slu.edu
>
>
>
> ------------------------------
>
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1014): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1014__;!!PvDODwlR4mBZyAb0!WJ78YHXkN4DWs9Ge1hSQmtKyHH8YAU=
BGctF4M6O5QwKnA9jZJSzYyk408eAIufn-CY7FlzhbuQbYmA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104533230=
/7559972__;!!PvDODwlR4mBZyAb0!WJ78YHXkN4DWs9Ge1hSQmtKyHH8YAUBGctF4M6O5QwKnA=
9jZJSzYyk408eAIufn-CY7Flzj4rPRe3A$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Sat Feb 24 2024 - 06:02:37 MST