[AMMRL] SampleXpress misbehaving

From: ter Horst, Marc A. <terhorst_at_email.unc.edu>
Date: Mon, 5 Feb 2024 20:52:48 +0000

Hello

   We are having issues with sample eject and sample down detection on our
60 position SampleXpress on a 600 MHz Ultrashield magnet with a AVIII spectrometer.
The two biggest issues: (1) Topspin 3.6.5 does not detect a sample down even
when a sample is in the magnet/probe and the SampleXpress detects a sample
down and (2) sample eject can fail intermittently as well as physically 'lock'
the spinner in the slot on the cassette. This happened after re-installing
the SampleXpress (and cryoprobe) after removing and cleaning the BST (it was
dirty as was the sensor). We have also added a high pressure N2 tank to
ensure proper source pressures.

A couple things we have tried:

* Power cycling the SampleXpress will bring operation back for 1 or
a few successful 'sx' commands but will eventually fail.

* All tubing is connected and pressures within normal ranges

* Sample eject adjust on the SampleXpress has been calibrated multiple times

* In the service mode, different voltage thresholds have been tried as well
as 'autocalibrate' with no improvement

* Icon can work for hours despite TS not indicating sample down. Within
12-ish hours there will be a sample eject error

* 'cf' has been run, settings within 'ha' have been confirmed.

* AND the system works correctly with a different SampleXpress, borrowed
from another spectrometer.

Of course the last point seems to be the key piece of evidence! We have wondered
if the cause is a mistake we made re-installing the SampleXpress. Using the second
SampleXpress seems to indicate we are at least on the right track and does not
indicate errors on installation.

Bruker has been helpful troubleshooting- thank you Robert (and Igor)!

Before calling in a service engineer and possibly swapping components/boards on
the malfunctioning SampleXpress, any similar experiences with possible solutions?
Marc



My work day may differ from your work day. Please do not feel obligated to
respond outside your working hours.
=-=-=-=-=-=-=
Marc ter Horst (he/him/his)
Research Professor
NMR Laboratory

056 Caudill Labs
Department of Chemistry, CB #3290
UNC at Chapel Hill
Chapel Hill, NC 27599-3290

e-mail: terhorst_at_unc.edu
NMR Lab: (919) 962-1149
Office: (919) 843-5802



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#975): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/975__;!!PvDODwlR4mBZyAb0!Ssc_19MjcN1GM5uP3lFosALkw1v_27PI=
e4ov7NBTawzcmuJOBCDSFYekq0b3SDIsFOnJAFzdO0SlMq1t3Auc2A$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104185349=
/7559972__;!!PvDODwlR4mBZyAb0!Ssc_19MjcN1GM5uP3lFosALkw1v_27PIe4ov7NBTawzcm=
uJOBCDSFYekq0b3SDIsFOnJAFzdO0SlMq0n8IIlIQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Feb 05 2024 - 12:53:19 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 29 2024 - 14:59:21 MST