The fix I found that seems most likely to resolve the issue is:
Rename or remove the folder C:\Users\<current_user>\.topspin1\ and restart topspin.
This does sound like a software issue and most likely not hardware.
Cheers,Mike
On Monday, December 9, 2024 at 01:37:36 PM MST, Baleja, James D via groups.io wrote:
Hi all,
I’m getting an error message “Index: 1, Size: 1” when
trying to start up Topspin 3.5 (pl 6, but I get the same error message with
pl5 or pl 7) on our 600. (AVANCE III HD NMR SPECT. 600)
Prior to that our instrument was “limping” a bit, with various
errors message and the P90 jumping from the usual 7.5 usec to close to 24 usec.
Linewidth was superb though.
Any ideas?
Thanks for listening,
Jim
--