From: Ulrich via groups.io <Ulrich.Haunz=uni-konstanz.de_at_groups.io>
Date: Mon, 09 Dec 2024 17:35:11 +0100

Hello all,

what kind of spinners are you using?

The blue ones - these are quite light. Do you use also Young tubes? These
make it difficult. These tubes are quite heavy.
So the air pressure can be a bit too low.
There is only one option to change the air pressure.

ha -> BSMS -> there is the option to install and change the main parameters
 - also the flow for the lift up and down.

I use not the sample case, when I am running an young tube. I do it manually - lift up and down.
If the up is not working properly - I close the main tube of the magnet with my hand and release it fast open.
So this helps a bit, to get the stucked sample out.

Have a nice try.

Yours sincerely,

Ulrich Haunz

Am Montag, Dezember 09, 2024 16:45 CET, schrieb "Beth Moscato via groups.io":

> Hi all –
>
> We have a sample that is not ejecting from a Sample Case. The eject air is
> working, but it’s not popping all the way out. At the moment, it’s
> unclear if this is because the sample is too heavy or somehow damaged or if
> there’s an issue with the airflow. Either way, I suspect it’s going
> to be ejected if I can increase the airflow rate.
>
> With the Sample Xpress and XpressLite, the sample ejection airflow can be
> tweaked with a dial behind the device. With the Sample Case, there doesn’t
> seem to be a similar arrangement.
>
> As best as I can tell from the manual, the ejection air from the Sample Case
> is controlled by a series of screws (?). It doesn’t seem like there’s
> a built in ability to adjust the eject air at all.
>
> Any advice?
> Beth Moscato
>

--
=========================

Ulrich Haunz, Dipl.-Ing. (FH)
NMR-Spektroskopie

Raum L 710, L 807
NMR-Center M520, M521, M527
Postfach M 715
Fachbereich Chemie
Universitaet Konstanz

78457 Konstanz

Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
Tel.: +49 (0)170 9153213
FAX : +49 (0)7531 88-3898
=========================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1920): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1920__;!!PvDODwlR4mBZyAb0!TbOLNRaQhS3WPZgeJw9vr2MoH7q77p=
-5SXzp0Aba69TZ5g9bgglfm9lnM07RD8XBW-yUxWRsZq2t7ZW5WATEozcNL9Pv$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/110007915=
/7559972__;!!PvDODwlR4mBZyAb0!TbOLNRaQhS3WPZgeJw9vr2MoH7q77p-5SXzp0Aba69TZ5=
g9bgglfm9lnM07RD8XBW-yUxWRsZq2t7ZW5WATEo3xb4xJv$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Mon Dec 09 2024 - 08:35:43 MST

This archive was generated by hypermail 2.4.0 : Sun Jan 05 2025 - 13:37:38 MST