Re: [AMMRL] Topspin acquisition license expired (?)

From: Bob Berno via groups.io <bberno=mcmaster.ca_at_groups.io>
Date: Tue, 18 Aug 2026 10:37:37 -0400

Hello,

I encountered something like this once on a completely different system
running a completely different version of TopSpin.
In my case, the "simple fix" was to completely reboot the workstation.

In my case, I think the issue was that the FlexLM service had stopped
running for some reason.  The reboot restarted license manager and
everything was fine again.  I know there are other ways to restart
FlexLM without resorting to a system reboot, but you asked about a
simple fix.

If FlexLM is running, but you still get the "No license..." error
message, then maybe your firewall rules have been updated, thus
preventing the system to work.

Good luck,

Bob.

On 2026-08-17 4:40 p.m., William Jarrett via groups.io wrote:

>
> Colleagues:

> This concerns an Avance III 600 install in 2012 and using Topspin
> 3.1pl7. I was help a user set up a VT NMR acquistion when I got the
> following error message:
>
> No license for TOPSPIN acquisition
>
> Examination of the license file indeed indicate that although every
> other module was dated 13-jun-2028, the acquitision was dated
> 29-jul-2026. This is especially frustrating as the expiration warning
> issued when starting Topspin only stated the 2028 timelimit. Does
> anyone know if there is a simple fix, or am I in for a long semester?
>
>
> Regards,
> WLJ
>
>
> --
> Dr. W. L. Jarrett
> University of Southern Mississippi
> School of Polymer Science and Engineering
> Phone: (601)266-4551
>
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One
Spoon" wampum agreement.





-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3699): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3699__;!!PvDODwlR4mBZyAb0!WdfSkVbLfR5myR0MepsYVrhhlRtC1c=
-NQDEbRmXe2uabIXWfYLfs78PjubXdnCfHNTiHS1HzeQ0V_O2bUbhjXRqIq_90$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120806311=
/7559972__;!!PvDODwlR4mBZyAb0!WdfSkVbLfR5myR0MepsYVrhhlRtC1c-NQDEbRmXe2uabI=
XWfYLfs78PjubXdnCfHNTiHS1HzeQ0V_O2bUbhjXYGinlis$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!WdfSkVbLfR5myR0Me=

Received on Tue Aug 18 2026 - 07:52:16 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST