Hi AMMRL Community.
Thanks so much for your help. I received several insightful responses, and
I think I’ve got a handle on the situation now.
Recap of the problem: My SampleCasePlus (60 samples) has been stopping the
queue because it times out when moving laterally between concentric rings of
samples. Cleaning its aluminum “fins” on top has helped in
the past, but the problem comes back intermittently, and incidents increase
in frequency over time. The unit’s manual doesn’t have an
entry for regular maintenance of the lower unit of autosampler.
Brief summary, with insights from several respondents:
*
This appears to be a common issue with the SampleCasePlus.
*
The usual treatment is occasionally wiping off the aluminum “fins”/plates
with a Kimwipe.
*
Several people have had success by removing the grease from the aluminum fins
that is present when the unit is first unpacked.
*
Contrary to my earlier speculations, it is not important for the fins/plates
to stick to one another to form a seal over the SampleCasePlus machinery.
This increases resistance to sliding and the grease attracts dust.
*
My original strategy was to clean the fin surfaces with a kimwipe and remove
the surface grease by spraying with 70% isopropanol and wiping it off, but
leaving the apparent white grease in the “rails” at the sides
(that keep the individual plates stacked up).
*
That method doesn’t look like the best solution. I think the grease
at the sides eventually works its way on to the plates, where it attracts
dust, and the problem reoccurs.
*
One respondent reported success by spraying WD-40 on the fins, which lubricates
them conveniently without removing the lower unit’s housing, and
doesn’t appear to be attractive to dust.
*
My personal favorite recommendation is to use some “Non-Stick Dry Film
Lubricant” from DuPont; it’s even got a little hexagon with
“BN” inside to indicate it’s based on boron nitride.
(
https://tinyurl.com/3mu7ps2t ) The form of the recommended product is a
4oz drip bottle, but I can only find it in spray cans, so I’ll try
the spray.
*
The respondent reported success with applying it to all flat surfaces of the
fins/plates, not just the “rails” at the edges. They and I
have both observed scratches, shiny spots, and other marks indicating wear
in certain places, so it seems like a sturdy, non-sticky lubricant would help.
*
One respondent suggested checking the alignment of the housing with the alignment
bolt at the bottom of the housing. Misalignment could cause the housing to
pivot at an unintended able, perhaps giving rise to scraping of the white
barrel under the top of the housing.
In sum, my plan is now to get a can of the DuPont BN-based Non-Stick Dry Film
Lubricant spray unless I can find a drip bottle. Once I have that, we’ll
disassemble the aluminum fins, clean them entirely and remove all the grease,
then re-lubricate the rails and flat surfaces with the new dry ubricant.
Fingers crossed!
Many thanks to Eugenio, Gabriel, Paris, and those who replied privately for
all their insights and suggestions!
Below please find some ancillary analysis concerning the relationship of the
mystery white powder to this problem, and identification of surfaces that
probably need lubrication on a regular basis.
Thanks so much, AMMRL community!
*
Josh
OTHER OBSERVATIONS - EFFECTS ON THE WHITE BARREL
Last week I removed the lower unit’s housing to clean the fins and
grease the rails on which the carousel slides back and forth, and also to
investigate the white powder that’s been accumulating on the aluminum
fins.
[1_2026-0811_SampleCasePlus_09_WhitePowderOnFins.jpg]
I checked whether the top edge of the white barrel might be scraping against
the underside of the housing. There as a definite gap between them, so there
was no scraping.
[2_2026-0811_SampleCasePlus_10_GapBarrel-Housing.jpg]
I checked alignment of the housing with the alignment bolt at the bottom. The
“top” of the bolt was flush with the surface of the housing.
I understand this is the proper position, since the bolt head fits precisely
inside and such placement secures the housing against pivoting around the
support pins at the top of the housing. If it’s screwed in too much
further in or out, it doesn’t perform this function.
[3_2026-0811_SampleCasePlus_12_AlignmentBolt.jpg]
Inspection of the top of the plastic barrel revealed significant wear. Grooves
were apparent in the barrel’s vertical surface that contacts the
metal fins. This strongly suggests that the resistance force provided by the
sticky fins has been strong enough that plastic has been getting scraped off
when the barrel rotates.
[4_2026-0814_SampleCasePlus_15_GroovesOnWhiteBarrel.jpg]
I am hoping that a regimen of regular cleaning and dry-film lubrication will
prevent the barrel from getting grooved-out like this in the future.
OTHER SURFACES TO LUBRICATE
So far, I’ve normally lubricated the easily-seen rails at the left-hand
side (when facing the magnet), along which the carousel “carriage”
slides. When one shuts off the pneumatic gas supply and pushes/pulls on the carousel,
it’s obvious these are being used for guidance and support. This time, I
did some more searching and found additional surfaces that appear to also support
and guide the carriage, plus the piston (“actor” in the manual) that
pushes and pulls the assembly back and forth. That piston shaft felt dry, as did
the other rails, so applied some Festo grease to them, probably for the first time
since installation about three years ago.
[5_2026-0814_SampleCasePlus_20_PistonToBeGreased.jpg]
The system has been running with no autosampler errors for several days.
P.S. I still don’t know what FSM means in this context, and I continue
to recognize it as “Flying Spaghetti Monster”.
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Josh Kurutz via groups.io
> Date: Wednesday, August 5, 2026 at 12:17 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] AMMRL: Intermittent SampleCasePlus (60) problem - stuck in state mid-to-in, mid-to-out, etc.
Hi AMMRL Community.
I'd like your help addressing a recurring intermittent problem with our
SampleCasePlus (60 samples). It occasionally times out as the carousel slides
back and forth to change which of the three rings of samples will be under
the transport tube. We get a “red bar” error, a failed sample,
and the queue stops. The error message is along the lines:
“…Fatal Error…1000 STOP_CHANGER…Timeout while
trying to move carousel. Move tray FSM is stuck in state 9: moving mid to in.”
Similar errors appear for moving between other ring positions (mid, in, out).
Recovery by resetting the autosampler with the emergency stop button is = immediate -
there’s no apparent change in carousel position upon successful reset -
like the carousel was just slow or needed another couple seconds to read its position.
I’d like some help either fixing the recurring problem or managing
my expectations. I’ve been focusing on the mechanical issues, but
perhaps there’s a setting in the HA web interface where I can increase
the timeout time by a couple seconds.
In the past, we’ve gained some relief by:
*
Cleaning the aluminum fins that cover the lower unit’s interior at
the front and back margins. Last time we did this, we removed the fin assembly
from the cover (but kept it intact) so we could place it flat on a clean bench
top, and wiped down the fins repeatedly with 70%-isopropanol-laden paper towels.
We took care to not disturb the side edges, which are apparently lubricated
with a white grease.
*
Using dry kimwipes to wipe down the interior track along which the carousel
(“tray”?) slides back and forth between its three positions.
I then applied a light coat of Festo grease (the same as used to lubricate
the slider rail atop the magnet) to that track.
*
Gently swabbing all sensors with cotton-tipped swabs briefly dipped in
electronics-grade isopropanol, paying extra attention to the ones sensing
mid/in/out positions.
That cleaning helped a lot a few months ago, but over the last few weeks the
problems have been occurring again, and incident frequency is increasing.
I also noticed some white powder occasionally builds up on top of one part
of a fin, right next to where the white carousel plastic protrudes above the
aluminum. This powder doesn’t really look like a spilled sample -
my suspicion is that something is rubbing or scraping the carousel plastic
and leaving a residue, but I don’t have a good basis for thinking
that, and I’d be hard-pressed to identify where that action might
happen.
The portion of the SampleCasePlus manual covering regular maintenance doesn’t
appear say anything about maintaining the fins or lubricating the carousel’s
sliding track inside the lower unit.
I’d asked Bruker’s Rover support about this before, and I got
some good insight. When I first got the units, the aluminum fins were lightly
greased, so I thought I should occasionally wipe them down and re-grease them.
An engineer corrected me on that, saying the grease doesn’t really
have a function when the unit is working, so I should just remove the grease.
However, I still figured the white grease at the sides looks important to
reduce wear.
Any ideas? Should I just start scheduling disassembly and cleaning every few
months?
Thanks. - Josh
P.S. What does the “FSM” in the error message mean, with respect
to the tray? I normally read this acronym with an entirely nontechnical meaning.
--
Josh Kurutz, PhD
NMR Facility Manager, Chemistry Dept.
University of Chicago
https://chemnmrlab.uchicago.edu/
jkurutz_at_uchicago.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3700): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3700__;!!PvDODwlR4mBZyAb0!T6J13g9OdOEWu9lTxfMKl8cQZIjshK=
uxuHOltM8JsRuxhJnO8-eJ2glBYKk2t4ezcnvfDlOo9DIMuw1ggyzfCbsoRTrz$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120613914=
/7559972__;!!PvDODwlR4mBZyAb0!T6J13g9OdOEWu9lTxfMKl8cQZIjshKuxuHOltM8JsRuxh=
JnO8-eJ2glBYKk2t4ezcnvfDlOo9DIMuw1ggyzfCXj6ubsR$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Aug 18 2026 - 10:09:44 MST