[AMMRL] SampleCase fails to load samples #Bruker #Hardware #NMR_Liquids

From: Piotr Dobrowolski via groups.io <piotr.dobrowolski=ttu.edu_at_groups.io>
Date: Wed, 05 Aug 2026 09:49:50 -0700

400 MHz Avance III HD / Topspin 3.6.5
We observe the following three different, intermittent behaviors of the SampleCase:
*
The sample loads normally, the lights on the changer are steady yellow

*
The sample loads into the magnet and immediately ejects, the system tries to
load several times and aborts, returning the sample to the carousel. Error
message: "Error STOP_CHANGER 1000 STOP_CHANGER Sample Transporter. Insert
carrier failed! Check carrier transition from SampleMail to BST. If Spin i s
working correctly and if VT gas flow is not too high. Try to increase the
BST Spinner Down Timeout (ServiceWeb)." This error stops automation run
making IconNMR useless.

*
The sample gets stuck at the top of the BST and does not load into the magnet,
but the lights on the changer are YELLOW! We can also see the turbine in the
gap between the transporter shuttle and the BST.Can this problem be related
to some random air pressure/flow issues, bad sensors, or something else? I
am a new Bruker user and I am absolutely confused.

Piotr


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3656): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3656__;!!PvDODwlR4mBZyAb0!SKBw3NI4nrHu0BbhldOpIwW8MMBCp-=
KpTwRycZsDaW9J_uk2xqdDkTPgqHkhYnZYHiKKSp_rIqdPp77gky7obdsiBJOQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120613924=
/7559972__;!!PvDODwlR4mBZyAb0!SKBw3NI4nrHu0BbhldOpIwW8MMBCp-KpTwRycZsDaW9J_=
uk2xqdDkTPgqHkhYnZYHiKKSp_rIqdPp77gky7obWMGyM_7$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!SKBw3NI4nrHu0BbhldOpIwW8MMBCp-KpTwRycZ=
sDaW9J_uk2xqdDkTPgqHkhYnZYHiKKSp_rIqdPp77gky7obY3xWm1R$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!SKBw3NI4nrHu0BbhldOpIwW8MMBCp-KpTw=
RycZsDaW9J_uk2xqdDkTPgqHkhYnZYHiKKSp_rIqdPp77gky7obUIy0mGg$
Mute #nmr_liquids:https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/nmr_liquids__;!!PvDODwlR4mBZyAb0!SKBw3NI4nrHu0BbhldOpIwW8MMBC=
p-KpTwRycZsDaW9J_uk2xqdDkTPgqHkhYnZYHiKKSp_rIqdPp77gky7obcYppKv-$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 05 2026 - 10:18:07 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST