Hi Piotr,
A common cause for this intermittent issue is the optical sensor at the top
of the magnet failing to detect the black strips on the spinner.
If there are minor fluctuations in the drive/lift airflow, the spinner can
sit slightly out of range of the sensor's light path. You can test this by
gradually reducing the airflow in small increments (around 2 psi at a time)
to see if detection improves. Ultimately, getting the airflow precisely
calibrated and stabilized should resolve the improper ejection/loading
errors.
Regards,
Srinivasa Penumutchu
University of Pennsylvania
Biophysics Department
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3661): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3661__;!!PvDODwlR4mBZyAb0!Xhma-sxcMaM_RupdBLz22MUWehcQK4=
-sxJVh7jYcIfeiDnO-ti3N63xNDej66mgo5a8hOqq0rDZIMJIjyb6tDmbSOOFi$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120613924=
/7559972__;!!PvDODwlR4mBZyAb0!Xhma-sxcMaM_RupdBLz22MUWehcQK4-sxJVh7jYcIfeiD=
nO-ti3N63xNDej66mgo5a8hOqq0rDZIMJIjyb6tDpxHbiKv$
Mute #bruker:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!Xhma-sxcMaM_RupdBLz22MUWehcQK4-sxJVh7j=
YcIfeiDnO-ti3N63xNDej66mgo5a8hOqq0rDZIMJIjyb6tDkAs6Cgi$
Mute #hardware:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!Xhma-sxcMaM_RupdBLz22MUWehcQK4-sxJ=
Vh7jYcIfeiDnO-ti3N63xNDej66mgo5a8hOqq0rDZIMJIjyb6tDjhpAKLB$
Mute #nmr_liquids:
https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/nmr_liquids__;!!PvDODwlR4mBZyAb0!Xhma-sxcMaM_RupdBLz22MUWehcQ=
K4-sxJVh7jYcIfeiDnO-ti3N63xNDej66mgo5a8hOqq0rDZIMJIjyb6tDm366vWN$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Aug 05 2026 - 12:12:12 MST