[AMMRL] RESOLVED: Diverging behavior detected during shimming (Profile changed)

From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Wed, 5 Aug 2026 19:19:37 +0000

Hi all,

Following up on my earlier post: this turned out to be a known intermittent
issue with the diffusion cabinet not fully handing off when switching between
the high-power diffusion amp (with preemphasis) and the GAB/3 for solutions work.

Fix: Disconnect the Ethernet to the diffusion cabinet, run cf, and it seems to
detect the GAB/3 normally.

Hope this saves someone a troubleshooting session.

Best,

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.