Re: [External Email] Re: [AMMRL] AMMRL: Accounting by computer login - How to prohibit gaming the system

From: Juergen Schulte via groups.io <schulte=binghamton.edu_at_groups.io>
Date: Tue, 26 Aug 2025 09:00:33 -0400

In Topspin versions 4.x each completed experiment has an “acqus”
file for the experiment status with the lines

##$DATE= 1755917610

##$DATE_START= 1755917468

The difference will be the duration in seconds.

Unfortunately, in versions 3.x the DATE_START is not included.

Best Regards,

Jürgen

_______________________________________________

Jürgen Schulte, Ph. D.
NMR Specialist
Binghamton University
SN Building, RM 1024
25 Murray Hill Road
Vestal, NY 13850

Ph.: 607-777-4405
Fax: 607-777-4478
schulte_at_binghamton.edu
<http://nmr.binghamton.edu/Schulte.html>


> From: main_at_ammrl.groups.io On Behalf Of Dave Rice via groups.io
> Sent: Monday, August 25, 2025 8:13 PM
> To: main_at_ammrl.groups.io
> Subject: [External Email] Re: [AMMRL] AMMRL: Accounting by computer login - How to prohibit gaming the system

Hi all

I share Margaret's interest in a way to monitor overall Bruker usage for manual
operation. I ask users to record their start and stop times in a paper log!
One thing I do, do for the less trustworthy users is to monitor the "audita.txt files"
in each data set. These files monitor key strokes and most important show a unique
signature for the start and stop of each aquisition. A "grep" to a unique string
in the file can pull out each acquisition start and stop for dates of an individual
month from the user's data directory.

Still I'd certainly be interested to know if there is a more general file
logging all console communication - similar to acctLog.xml on the Varians. If
somone was interested in describing Bruker logging I'd certainly be happy
for the information.

Dave


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2748): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2748__;!!PvDODwlR4mBZyAb0!QxSvz0bO_rANtghBw3gJx01UwJ_oJ3=
EZZX4PowIfTK0cpDyjc0ienR_fVof3fJtRVu9kYPjKnsOHwA8CJPe5aQ5LzCWV$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114899418=
/7559972__;!!PvDODwlR4mBZyAb0!QxSvz0bO_rANtghBw3gJx01UwJ_oJ3EZZX4PowIfTK0cp=
Dyjc0ienR_fVof3fJtRVu9kYPjKnsOHwA8CJPe5aaaMMtAx$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Aug 26 2025 - 06:16:57 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST