The workaround for this problem is to set the PC system clock back to 2020,
linux command:
date +%Y%m%d -s "20200817"
The solution is to upgrade. Thanks to all of you who responded so quickly.
Todd
On Thu, Aug 18, 2022 at 3:17 PM Todd Rappe <rapp0006_at_umn.edu> wrote:
> We have two older systems running Topspin 2.8 (CentOS 5). On both of them,
> using the atmm command to tune manually results in an error "Invalid state
> at begin_invocation".
>
> The wobble curve appears as normal, but all of the control buttons to
> adjust the tuning and matching are grayed out. Automatic tuning with atma
> still works fine.
>
> This behavior started on both systems at roughly the same time, which may
> just be coincidence. Has anybody encountered this problem and found a
> solution?
>
> Thanks,
> Todd
> _____________________________________
> Todd Rappe
> Minnesota NMR Center <nmr.umn.edu>
> 612-625-9147 (office)
> 651-321-TODD (rings office and mobile)
> 612-624-8892 (magnet lab)
>
Received on Thu Aug 18 2022 - 11:30:11 MST