Re: [AMMRL] FID spikes while running CPMAS

From: Hsin Wang <hwang1_at_ccny.cuny.edu>
Date: Tue, 23 Apr 2024 12:50:00 -0700

Vernon,

I have also seen frequent sporadic spikes in my fids of CP experiments, 
but only in samples containing carbon black. The spikes are not sufficient to
cause overflow warnings.  Sometimes I see ACB errors during the run. 
The ACB error goes away when experiment ended and spikes do not appear when I
resume running other materials. And of course the CP results are very noisy.
I have been told my probe should be able to handle carbon black.

Not sure if our observations are related.  But I find Jochem's reply
interesting and tantalizing.  If something electronic response is happening
in the rotor, how can this affect the oxygen outside and cause the coil to produce
spikes?   I would imagine using N2 to spin the sample is costly, and
it would be difficult to test this idea.

I did try to use less power for CP.  But I have not tried lowering my
TPPM decoupling.  I think I should try that.

Hsin


-=-=-=-=-=-=-=-=-=-=-=-