Re: [AMMRL] BLARH100 with Tecmag Redstone

From: Huadong Zeng <hzeng_at_mbi.ufl.edu>
Date: Wed, 17 Apr 2024 19:53:24 +0000

I used Bruker RF amplifiers (specifically BLAH1000 and BLAX1000) on both Agilent
and Bruker consoles before. There is a TTL jumper inside the amplifier to select
TTL low or high. Bruker console uses TTL active low and Agilent is just the opposite.

Best regards

Huadong Zeng, Ph.D.
Preclinical MRI/S
Advanced Magnetic Resonance Imaging and Spectroscopy Facility
National High Magnetic Field Laboratory
McKnight Brain Institute
University of Florida
1149 Newell Dr, LG-126
Gainesville, FL 32611
Office: (352)294-8638

> From: main_at_ammrl.groups.io On Behalf Of Spin-Doc via groups.io
>Sent: Tuesday, April 16, 2024 5:45 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] BLARH100 with Tecmag Redstone

Hi,

from a BLARH100 900 MHz User manual:

* Safety features (OVERDRIVE, PULSE WIDTH, DUTY CYCLE) which will temporarily
disable the amplifier whenever output power is above a specified limit. Each
amplifier has on board information stored in the RS485 Interface Board regarding
amplifier type, max. Duty Cycle, max. Pulse Width etc. These parameters can be
changed by software if necessary using the 'edacb' command.

* The amplifiers are controlled by the ACB-X (Amplifier Control Board Extended)
via an RS485 type link (SBSB2 / tty20 Bus). Individual amplifiers can be addressed.
Hardware address codes are set via a HEX. switch located on the amplifier front panel.
Read and write access to the BLAD Control board is possible via the ACB-X Board.

* Precise software timing control of the amplifier blanking pulses (BLKTR1..15) via the "edscon" table.
The blanking pulses (BLKTR 1... 15) are active low TTL level and generated
by the SGU. They are transmitted to the PSD via the backplane and then to the
amplifier via external cabling. They are easily measured at the BNC connections
on the amplifier front panel. Which blanking pulse is required for which amplifier
will depend upon which router output provides the RF amplifier input. Thus an
amplifier channel connected to router output RO1 will be blanked by BLKTR1. An
amplifier channel connected to router output RO3 will be blanked by BLKTR3 etc.
The source of an individual BLKTR signal will be the SGU on which the corresponding
RF signal has been generated. This ensures coherence between the SGU blanking and
the external amplifier blanking. The timing of the BLKTR[1... 15] pulses can be
adjusted with the "edscon"= table.

Best regards
Bert


Am 16.04.2024 um 20:59 schrieb Joseph Sachleben:

Hi All,

I'm trying to control a an old Bruker BLARH100 500-600 MHz Amplifier with my
Tecmag Redstone. I can't find a manual for it however. I found the manual
that is in the Bash documentation that shows the block diagrams of the circuits,
but I can't find a users manual that tells me if it is using positive or negative
logic for the blanking. What the difference is between the high, medium, and low
inputs and outputs on the amp? What's the function of the RS486 port is?

Any help with documentation would be greatly useful.

Thanks,

Joe S

Joseph R. Sachleben, Ph. D.
Technical Director Biomolecular NMR Facility
The Division of Biological Sciences
929 E. 57th St.
Chicago, IL 60637

(773)834-9866
jsachleben_at_uchicago.edu


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)

Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net<mailto:bert.heise_at_spin-doc.net>
Web:
http://www.spin-doc.ne

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1161): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1161__;!!PvDODwlR4mBZyAb0!Qrkb8g010F-fS5lDqyHpLJ6T-FOt4N=
QQQ0qnsR0MRTMWkbnXpaiXTBQjygGrkwSupuk8r1UxUiE9i6Ed$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105563537=
/7559972__;!!PvDODwlR4mBZyAb0!Qrkb8g010F-fS5lDqyHpLJ6T-FOt4NQQQ0qnsR0MRTMWk=
bnXpaiXTBQjygGrkwSupuk8r1UxUkJDjuBD$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Apr 17 2024 - 12:59:44 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST