From: Joseph Sachleben <jsachleben_at_uchicago.edu>
Date: Wed, 17 Apr 2024 21:18:05 +0000

Thanks Huadong.

Do you know if you need to set whether it uses the high, med, low portions of the amp?
Right now, I’m not getting much amplification at all. I don’t know if
this old amp is bad, or if I’m not controlling it properly.

Joe

On Apr 17, 2024, at 2:53 PM, Huadong Zeng <hzeng_at_mbi.ufl.edu> wrote:

I used Bruker RF amplifiers (specifically BLAH1000 and BLAX1000) on both Agilent
and Bruker consoles before. There is a TTL jumper inside the amplifier to select
TTL low or high. Bruker console uses TTL active low and Agilent is just the opposite.

Best regards