Re: [AMMRL] Chlorine NMR for trace compounds in bulk Cl- electrolyte?

From: Mark Swanson <mark94131_at_gmail.com>
Date: Tue, 9 Apr 2024 17:09:25 -0700

Hi Wendy,

Both Chlorine isotopes are quadrupolar which will give you broad signals.
It looks like frequency is pretty low too so your probe may not be able to
get down that low but it's higher than 15N so it could be doable. You will
want to use a broadband probe if you have one but it could be worth a try.
The relaxation times are very short so you'll want to pulse rapidly and
might need several hours. I'm concerned about the MgCl2 being 3M as it
will swamp the other signals and I don't think you'll be able to see
dissolved Cl2 unless it's really soluble.

Here's some more info:

https://chem.ch.huji.ac.il/nmr/hebrew/techniques/1d/row3/cl.html

Good luck,

Mark


On Tue, Apr 9, 2024 at 4:47 PM Wendy Breyer <web216_at_lehigh.edu> wrote:

> Hi all,
>
> I'm looking for some advice on the feasibility of some proposed
> experiments. A user here would like to do Cl-NMR on aqueous electrolyte
> solutions. Specifically, he would like to check for small amounts of OCl-
> and dissolved Cl2 in the presence of 3M MgCl2. After searching the
> literature, I have been unable to locate any publications that include
> similar experiments so I suspect that his proposed experiments are not
> feasible... but my knowledge of NMR is superficial at best. I'd be very
> interested in hearing from those with more experience whether these
> experiments are possible. (For context, we have a 400 MHz Avance III HD
> with a room temperature, broadband probe.)
>
> Thanks for any input!
>
> Wendy
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1121): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1121__;!!PvDODwlR4mBZyAb0!XKV6a7t76H7LUvkD0AMVOsNhbRRqFm=
vvwpPXoCSUFqhZaUj2F7hqDNJVY5v8b8VYZIEHc_dyuyyLhZeA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105433086=
/7559972__;!!PvDODwlR4mBZyAb0!XKV6a7t76H7LUvkD0AMVOsNhbRRqFmvvwpPXoCSUFqhZa=
Uj2F7hqDNJVY5v8b8VYZIEHc_dyu0kYZFHu$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Apr 09 2024 - 17:09:45 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST