Hi Julian,
The power supply in our Agilent DD2 800 shim power supply failed a while back
which caused all the shims and Z0 to become unresponsive. When the problem
occurred all the internal LEDS in the shim power supply were out, confirming
the problem was with the power supply.
We sent the unit to Kevin Goehring at BlueSky NMR (
https://www.blueskynmr.com/)
for repair, and it has been working well since then.
Unfortunately as our Agilent and Varian spectrometers age we are seeing more
power supply problems.
Best regards,
Eugene
Eugene DeRose, Ph.D. - Contractor
NMR Core Manager
National Institute of Environmental Health Sciences (NIEHS)
National Institutes of Health (NIH)
Bldg. 101/MRI-32
984.287.3572
derose_at_niehs.nih.gov
--
> From: main_at_ammrl.groups.io on behalf of Julian Holzinger via groups.io
> Date: Friday, September 11, 2026 at 9:56 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [EXTERNAL] [AMMRL] Shim not reacting on Varian VNMR 400
Hi everyone,
we are having a problem with our Varian VNMR 400 system: recently, a fuse
in the 5V-channel of the main power supply blew, which we subsequently replaced.
Then we realised that we cannot lock anymore (see photo attached); however,
when changing lock phase and gain the lock signal responds, whereas it does
not when changing z0. Also when changing z/x/y-shims there is no change in
the lock signal, which made us believe that it has smth to do with the shims.
We checked all the cables, even swapped the shim power supply with another
one, and checked every output on the 5V line to see if there are actually
5V present (which was the case). We are able to acquire spectra (with a
horrible lineshape, of course) but still cannot lock/shim.
Has anyone of you faced similar problems and has any idea?
Best regards and a nice weekend
Julian
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3778): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3778__;!!PvDODwlR4mBZyAb0!QTI5ue_6NK6fGfA5e0rMDVt0iUPQWj=
_TPTacME-WQah8k10LAiPok9277WWpHrtcZYuY6VJ0dJonQGLF9lC1NFjFBd2n$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121198428=
/7559972__;!!PvDODwlR4mBZyAb0!QTI5ue_6NK6fGfA5e0rMDVt0iUPQWj_TPTacME-WQah8k=
10LAiPok9277WWpHrtcZYuY6VJ0dJonQGLF9lC1NOsGGCP5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Sep 11 2026 - 10:04:24 MST