Hi everyone,
we are having a problem with our Varian VNMR 400 system: recently, a fuse
in the 5V-channel of the main power supply blew, which we subsequently replaced.
Then we realised that we cannot lock anymore (see photo attached); however,
when changing lock phase and gain the lock signal responds, whereas it does
not when changing z0. Also when changing z/x/y-shims there is no change in
the lock signal, which made us believe that it has smth to do with the shims.
We checked all the cables, even swapped the shim power supply with another
one, and checked every output on the 5V line to see if there are actually
5V present (which was the case). We are able to acquire spectra (with a
horrible lineshape, of course) but still cannot lock/shim.
Has anyone of you faced similar problems and has any idea?
Best regards and a nice weekend
Julian
-=-=-=-=-=-=-=-=-=-=-=-