Re: [AMMRL] AU Program for Automated Sample Queuing in TopSpin?

From: Paris Ning, PhD via groups.io <pning=isotope.com_at_groups.io>
Date: Wed, 2 Sep 2026 11:58:02 +0000

Hi Benjamin,

Just wanted to know if the NMR without ICON is equipped with sample changer,
or if it simply doesn't have an ICON license.

If your goal is to simplify the commands a user needs to type in to run an
experiment, you can implement an AU with "xcmd" lines to call the commands
(eg. "xau atma") as it waits for the current commands to finish before
moving on to the next line.

Hope this helps.

Cheers,
Paris

> From: main_at_ammrl.groups.io On Behalf Of Atterberry, Benjamin via groups.io
> Sent: Tuesday, September 1, 2026 3:24 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] AU Program for Automated Sample Queuing in TopSpin?

Hi all,

I am the NMR Manager at Mississippi State University. Do you by any chance
have an AU program for Bruker TopSpin that runs samples one after the other,
like IconNMR does? What I'm looking for is something where you can create
different datasets in TopSpin, then queue up sample insertion, lock,
automatic tuning, topshim, getprosol, optimizing receiver gain, and zg.

We have one spectrometer that gets crowded because it has IconNMR, so the users
like it because they can queue up several 1H experiments. If we could use
an AU program to "impersonate" IconNMR on another spectrometer, it would
free up this one instrument.

Best,
Benjamin Atterberry

----------------------------------
Benjamin A. Atterberry
NMR Facility Manager
Department of Chemistry
Office: Hand Lab G052
Mississippi State University


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3739): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3739__;!!PvDODwlR4mBZyAb0!XgTMwGjZzsaSizadYa0vkY-4YHfvLp=
GMIyejKRlxVqb-32RGUOkSrNnYkXlFU6UmpbgobMubeWv-4B-BJ1l52yWAMRQI$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121039594=
/7559972__;!!PvDODwlR4mBZyAb0!XgTMwGjZzsaSizadYa0vkY-4YHfvLpGMIyejKRlxVqb-3=
2RGUOkSrNnYkXlFU6UmpbgobMubeWv-4B-BJ1l52zuegTSB$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Sep 02 2026 - 05:51:04 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST